Technical Note: Immunohistochemical evaluation of mouse brain irradiation targeting accuracy with 3D‐printed immobilization device
- 16 October 2015
- journal article
- radiation biology
- Published by Wiley in Medical Physics
- Vol. 42 (11) , 6507-6513
Abstract
Purpose: Small animal immobilization devices facilitate positioning of animals for reproducible imaging and accurate focal radiation therapy. In this study, the authors demonstrate the use of three‐dimensional (3D) printing technology to fabricate a custom‐designed mouse head restraint. The authors evaluate the accuracy of this device for the purpose of mouse brain irradiation.Methods: A mouse head holder was designed for a microCT couch using cad software and printed in an acrylic based material. Ten mice received half‐brain radiation while positioned in the 3D‐printed head holder. Animal placement was achieved using on‐board image guidance and computerized asymmetric collimators. To evaluate the precision of beam localization for half‐brain irradiation, mice were sacrificed approximately 30 min after treatment and brain sections were stained for γ‐H2AX, a marker for DNA breaks. The distance and angle of the γ‐H2AX radiation beam border to longitudinal fissure were measured on histological samples. Animals were monitored for any possible trauma from the device.Results: Visualization of the radiation beam on ex vivo brain sections with γ‐H2AX immunohistochemical staining showed a sharp radiation field within the tissue. Measurements showed a mean irradiation targeting error of 0.14 ± 0.09 mm (standard deviation). Rotation between the beam axis and mouse head was 1.2° ± 1.0° (standard deviation). The immobilization device was easily adjusted to accommodate different sizes of mice. No signs of trauma to the mice were observed from the use of tooth block and ear bars.Conclusions: The authors designed and built a novel 3D‐printed mouse head holder with many desired features for accurate and reproducible radiation targeting. The 3D printing technology was found to be practical and economical for producing a small animal imaging and radiation restraint device and allows for customization for study specific needs.Keywords
Funding Information
- London Regional Cancer Program Catalyst Grant (LHR-F8150)
- Translational Breast Cancer Studentship
- London Regional Cancer Program Catalyst Grant (LHR-F8150)
- Translational Breast Cancer Studentship
This publication has 15 references indexed in Scilit:
- Image-guided microbeam irradiation to brain tumour bearing mice using a carbon nanotube x-ray source arrayPhysics in Medicine & Biology, 2014
- Development and validation of a treatment planning system for small animal radiotherapy: SmART-PlanRadiotherapy and Oncology, 2013
- Implementation and commissioning of an integrated micro-CT/RT system with computerized independent jaw collimationMedical Physics, 2013
- Abstract 5725: Accuracy of off-line bioluminescence imaging to localize targets in pre-clinical radiation researchCancer Research, 2012
- Localized CT-Guided Irradiation Inhibits Neurogenesis in Specific Regions of the Adult Mouse BrainRadiation Research, 2011
- Characterization of image quality and image‐guidance performance of a preclinical microirradiatorMedical Physics, 2011
- CT Guidance is Needed to Achieve Reproducible Positioning of the Mouse Head for Repeat Precision Cranial IrradiationRadiation Research, 2010
- Feasibility of small animal cranial irradiation with the microRT systemMedical Physics, 2008
- Characteristics of γ-H2AX foci at DNA double-strand breaks sitesBiochemistry and Cell Biology, 2003
- Evidence for a lack of DNA double-strand break repair in human cells exposed to very low x-ray dosesProceedings of the National Academy of Sciences of the United States of America, 2003