A pH Sensor Based on a Stainless Steel Electrode Electrodeposited With Iridium Oxide
Authors:
- 17 November 2008
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Education
- Vol. 52 (1) , 133-136
Abstract
A simple procedure to make an iridium oxide (IrO2) electrodeposited pH sensor, that can be used in a chemical, biomedical, or materials laboratory, is presented here. Some exercises, based on this sensor, that can be used to teach important concepts in the field of biomedical, biochemical, tissue, or materials engineering, are also presented. This novel procedure is based on the electrodeposition of IrO2 on a stainless steel electrode, which uses a similar mechanism to an ion selective electrode (ISE) and senses changes in pH. The simplicity and cost effectiveness of this method facilitates the teaching of the concept of half-cell potential and the basics of sensors. This novel sensor has also been shown to outperform the classical glass pH-sensor. In this new methodology, students learn to build the electrode, to calibrate it, and to measure its sensitivity, repeatability, and time-response.Keywords
This publication has 11 references indexed in Scilit:
- Differential pH measurements of metabolic cellular activity in nl culture volumes using microfabricated iridium oxide electrodesBiosensors and Bioelectronics, 2007
- Metal–metal oxide pH sensors for physiological applicationMedical Engineering & Physics, 2006
- Thin-film IrO pH microelectrode for microfluidic-based microsystemsBiosensors and Bioelectronics, 2005
- Improved Electrodeposited Iridium Oxide pH Sensor Fabricated on Etched Titanium SubstratesAnalytical Chemistry, 2003
- Fabrication of Anodically Electrodeposited Iridium Oxide Film pH Microelectrodes for Microenvironmental StudiesAnalytical Chemistry, 2002
- Electrodeposited iridium oxide for neural stimulation and recording electrodesIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2001
- A pH Electrode Based on Melt-Oxidized Iridium OxideJournal of the Electrochemical Society, 2001
- Electrodeposited Iridium Oxide pH Electrode for Measurement of Extracellular Myocardial Acidosis during Acute IschemiaAnalytical Chemistry, 1998
- Chemical SensorsAnalytical Chemistry, 1998
- Laboratory scale electrodepositon: Practice and applicationsJournal of Chemical Education, 1986