Improved 3‐Tesla cardiac cine imaging using wideband
Open Access
- 4 May 2010
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 63 (6) , 1716-1722
Abstract
Cine balanced steady‐state free precession (SSFP) is the most widely used sequence for assessing cardiac ventricular function at 1.5 T because it provides high signal‐to‐noise ratio efficiency and strong contrast between myocardium and blood. At 3 T, the use of SSFP is limited by susceptibility‐induced off‐resonance, resulting in either banding artifacts or the need to use a short‐sequence pulse repetition time that limits the readout duration and hence the achievable spatial resolution. In this work, we apply wideband SSFP, a variant of SSFP that uses two alternating pulse repetition times to establish a steady state with wider band spacing in its frequency response and overcome the key limitations of SSFP. Prospectively gated cine two‐dimensional imaging with wideband SSFP is evaluated in healthy volunteers and compared to conventional balanced SSFP, using quantitative metrics and qualitative interpretation by experienced clinicians. We demonstrate that by trading off temporal resolution and signal‐to‐noise ratio efficiency, wideband SSFP mitigates banding artifacts and enables imaging with approximately 30% higher spatial resolution compared to conventional SSFP with the same effective band spacing. Magn Reson Med, 2010.Keywords
Funding Information
- National Institutes of Health (R21-HL079987)
- American Heart Association (#0435249N)
This publication has 16 references indexed in Scilit:
- Measurement and characterization of RF nonuniformity over the heart at 3T using body coil transmissionJournal of Magnetic Resonance Imaging, 2008
- Wideband SSFP: Alternating repetition time balanced steady state free precession with increased band spacingMagnetic Resonance in Medicine, 2007
- Cardiac Steady-State Free Precession CINE Magnetic Resonance Imaging at 3.0 TeslaInvestigative Radiology, 2006
- Alternating repetition time balanced steady state free precessionMagnetic Resonance in Medicine, 2006
- Artifact reduction in true‐FISP imaging of the coronary arteries by adjusting imaging frequencyMagnetic Resonance in Medicine, 2003
- Preliminary report on in vivo coronary MRA at 3 Tesla in humansMagnetic Resonance in Medicine, 2002
- Cardiac ejection fraction: Phantom study comparing cine MR imaging, radionuclide blood pool imaging, and ventriculographyJournal of Magnetic Resonance Imaging, 1992
- Variable-rate selective excitationJournal of Magnetic Resonance (1969), 1988
- An analysis of fast imaging sequences with steady‐state transverse magnetization refocusingMagnetic Resonance in Medicine, 1988
- Report of the American Society of Echocardiography Committee on Nomenclature and Standards in Two-dimensional Echocardiography.Circulation, 1980