Synthesis of Electrocardiogram V-Lead Signals From Limb-Lead Measurement Using R-Peak Aligned Generative Adversarial Network
- 21 August 2019
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Journal of Biomedical and Health Informatics
- Vol. 24 (5) , 1265-1275
Abstract
Recently, portable electrocardiogram (ECG) hardware devices have been developed using limb-lead measurements. However, portable ECGs provide insufficient ECG information because of limitations in the number of leads and measurement positions. Therefore, in this study, V-lead ECG signals were synthesized from limb leads using an R-peak aligned generative adversarial network (GAN). The data used the Physikalisch-Technische Bundesanstalt (PTB) dataset provided by PhysioNet. First, R-peak alignment was performed to maintain the physiological information of the ECG. Second, time domain ECG was converted to bi-dimensional space by ordered time-sequence embedding. Finally, the GAN was learned through the pairs between the modified limb II (MLII) lead and each chest (V) lead. The result showed that the mean structural similarity index (SSIM) was 0.92, and the mean error rate of the percent mean square difference (PRD) of the chest leads was 7.21%.Keywords
Funding Information
- National Research Foundation
- Ministry of Science, ICT and Future Planning (NRF-2018M3A9H6081483)
This publication has 24 references indexed in Scilit:
- ECG signal enhancement based on improved denoising auto-encoderEngineering Applications of Artificial Intelligence, 2016
- Frank vectorcardiographic system from standard 12 lead ECG: An effort to enhance cardiovascular diagnosisJournal of Electrocardiology, 2016
- Methods for derivation of orthogonal leads from 12-lead electrocardiogram: A reviewBiomedical Signal Processing and Control, 2015
- A Review of ECG Data Compression TechniquesInternational Journal of Computer Applications, 2015
- Accurate and reliable 3-lead to 12-lead ECG reconstruction methodology for remote health monitoring applicationsIRBM, 2014
- Derivation of the 12-lead electrocardiogram and 3-lead vectorcardiogramThe American Journal of Emergency Medicine, 2013
- Transformation of the Mason-Likar 12-lead electrocardiogram to the Frank vectorcardiogram2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), 2012
- A Novel Neural-Network Model for Deriving Standard 12-Lead ECGs From Serial Three-Lead ECGs: Application to Self-CareIEEE Transactions on Information Technology in Biomedicine, 2010
- Linear affine transformations between 3-lead (Frank XYZ leads) vectorcardiogram and 12-lead electrocardiogram signalsJournal of Electrocardiology, 2009
- PhysioBank, PhysioToolkit, and PhysioNetCirculation, 2000