The perfusion index measured by the pulse oximeter affects the agreement between ClearSight and the arterial catheter-based blood pressures: A prospective observational study
Open Access
- 10 July 2019
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 14 (7) , e0219511
Abstract
ClearSight is a noninvasive arterial blood pressure monitor, but it remains unknown whether it is affected by the state of perfusion to the fingers. We investigated whether the lower perfusion index (PI) measured with a pulse oximeter, which reflects finger perfusion, would affect the agreement between arterial pressures measured with ClearSight versus those obtained with an arterial catheter. Paired arterial pressure data (ClearSight and arterial catheter-based pressures) and PI values were prospectively obtained from 30 patients undergoing major abdominal surgery. The primary outcome was standard deviation (SD) of the bias (precision) of blood pressure between ClearSight and arterial catheter. The ratio of the adjusted SD of the bias between PI≤1 and PI>1 was calculated using the linear mixed-effects model. The secondary outcomes were the bias and the limits of agreement (LOA) between the two devices (repeated measures Bland-Altman analysis). We analyzed 6312 paired data points. The adjusted SD of bias in PI ≤1 compared with those in PI >1 was 1.4-fold (95% confidence interval: 1.3- to 1.4-fold) for systolic arterial pressure, 1.5-fold (95% confidence interval: 1.3- to 1.6-fold) for diastolic arterial pressure, and 1.3-fold (95% confidence interval: 1.2- to 1.5-fold) for mean arterial pressure. The bias (LOA) were as follows: systolic arterial pressure in the PI ≤1 and PI >1 groups, -3.5 (-35.4 to 28.4) mmHg and 2.2 (-19.9 to 24.3) mmHg, respectively; diastolic arterial pressure in the PI ≤1 and PI >1 groups, 13.1 (-5.1 to 31.3) mmHg and 9.0, (-2.6 to 20.6) mmHg, respectively; and mean arterial pressure in the PI ≤1 and PI >1 groups, 8.7 (-11.3 to 28.7) mmHg and 7.6 (-6.2 to 21.3) mmHg, respectively. PI ≤1 was associated with a large SD of the bias between the devices. The PI value could be a real-time indicator of ClearSight precision.Keywords
This publication has 21 references indexed in Scilit:
- Accuracy and Precision of Continuous Noninvasive Arterial Pressure Monitoring Compared with Invasive Arterial PressureAnesthesiology, 2014
- Non-invasive continuous arterial pressure monitoring with Nexfin® does not sufficiently replace invasive measurements in critically ill patientsBritish Journal of Anaesthesia, 2013
- Non-invasive continuous arterial pressure and cardiac index monitoring with Nexfin after cardiac surgeryBritish Journal of Anaesthesia, 2012
- Noninvasive Continuous Arterial Blood Pressure Monitoring with Nexfin®Anesthesiology, 2012
- Agreed StatisticsAnesthesiology, 2012
- Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurementBritish Journal of Anaesthesia, 2011
- Noninvasive cardiac output and blood pressure monitoring cannot replace an invasive monitoring system in critically ill patientsBMC Anesthesiology, 2009
- Agreement Between Methods of Measurement with Multiple Observations Per IndividualJournal of Biopharmaceutical Statistics, 2007
- Models of brachial to finger pulse wave distortion and pressure decrementCardiovascular Research, 1997
- Forming inferences about some intraclass correlation coefficients.Psychological Methods, 1996