Technical Note: Accuracy of MTF measurements with an edge phantom at megavoltage x‐ray energies
Authors:
Open Access
- 3 October 2019
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 46 (12) , 5685-5689
Abstract
Purpose Measurement of the modulation transfer function (MTF) is performed by evaluating the response of an imaging system to a predefined input. To obtain accurate results when using an edge phantom, the detector input signal must resemble an ideal step function. The MTF of megavoltage (MV) imagers used in radiotherapy has been measured with highly absorbing edge phantoms fabricated from thick metal blocks. This study investigates the influence of the edge phantom design on the accuracy of the resulting MTF. Methods The MTF of an electronic portal imaging device (EPID) was measured at 6 MV beam quality with four edge phantoms made of lead with 1.3 cm, 3.3 cm, 5.0 cm and 10.0 cm thickness. Monte Carlo simulations were carried out for these and a selection of tungsten phantoms to determine the photon fluence at the imaging plane and quantify the systematic error in the MTF introduced by the edge phantom design. Results The measured MTF depends on the design of the edge phantom. The detector input signal of a thin phantom is affected by secondary radiation from the phantom itself, causing an overestimation of the MTF. The amount of secondary radiation can be reduced by increasing the phantom thickness or introducing an air gap between the phantom and the detector. Both methods introduce geometric unsharpness, which can result in an underestimation of the true MTF. Edge phantoms made from 4.0 cm thick tungsten or 5.0 cm thick lead induce comparatively small systematic errors of below 3 % or 5 %, respectively. Conclusions When MTF measurements are conducted at MV energies, even a highly absorbing edge phantom will introduce a systematic error of several percent. Direct comparison of MTFs obtained with different edge phantoms should therefore be treated with caution.Keywords
This publication has 17 references indexed in Scilit:
- Modeling the TrueBeam linac using a CAD to Geant4 geometry implementation: Dose and IAEA‐compliant phase space calculationsMedical Physics, 2011
- Image Quality Metrics for Digital Systems2010
- Validity of the line‐pair bar‐pattern method in the measurement of the modulation transfer function (MTF) in megavoltage imagingMedical Physics, 2007
- Determination of the detective quantum efficiency of a prototype, megavoltage indirect detection, active matrix flat-panel imagerMedical Physics, 2001
- Characteristics of metal‐plate/film detectors at therapy energies. I. Modulation transfer functionMedical Physics, 1998
- Optimal radiographic magnification for portal imagingMedical Physics, 1994
- A simple method for determining the modulation transfer function in digital radiographyIEEE Transactions on Medical Imaging, 1992
- Therapy imaging: A signal‐to‐noise analysis of a fluoroscopic imaging system for radiotherapy localizationMedical Physics, 1990
- Therapy imaging: A signal-to-noise analysis of metal plate/film detectorsMedical Physics, 1987
- A megavoltage MTF measurement technique for metal screen‐film detectorsMedical Physics, 1979