In Vitro Evaluation of the Removal Force of Abutments in Frictional Dental Implants
Authors:
- João César ZielakJZJoão César Zielak
- Murilo RorbackerMRMurilo Rorbacker
- Rodrigo GomesRGRodrigo Gomes
- Celso YamashitaCYCelso Yamashita
- Carla Castiglia GonzagaCGCarla Castiglia Gonzaga
- Allan Fernando GiovanniAGAllan Fernando Giovanni
- 1 October 2011
- journal article
- Published by American Academy of Implant Dentistry in Journal of Oral Implantology
- Vol. 37 (5) , 519-523
Abstract
The objective of the present work was to determine some force parameters for removal of an abutment from a dental implant in a frictional system (locking taper, 1.23 degrees). Ten implants of the same length (11 mm) and different diameters were selected, along with 10 straight abutments (13 mm length) with different diameters. Abutments were attached to implants without application of force. Fixation of the implant-abutment mount (IA) (repeated 1–5 times) was performed through the impact of a body weight (compression force, tapping) left from a known height. After each group of tappings, IA mounts were coupled with a tensile strength tester. The lowest removal value was found after the first tapping of mount #2 (83 N, implant diameter 3.3 mm/4.5 mm abutment diameter), and the highest removal value happened with mount #8 after the fifth tapping (420 N, 5.0 mm/5.5 mm). The force to remove IA mounts increased with the number of tappings and with the increase in abutment mass. Three activations (tappings) of the abutment were considered necessary to yield optimal stability, demonstrated by the large increase in removal force.Keywords
This publication has 5 references indexed in Scilit:
- Oral implants – the futureAustralian Dental Journal, 2008
- The Integrated Abutment Crown™, a Screwless and Cementless Restoration for Single‐Tooth Implants: A Report on a New TechniqueJournal of Prosthodontics, 2007
- Efficiency Considerations for the Purely Tapered Interference Fit (TIF) Abutments Used in Dental ImplantsJournal of Biomechanical Engineering, 2004
- THE LOCKING TAPER ATTACHMENT FOR IMPLANT ABUTMENTS: USE AND RELIABILITYImplant Dentistry, 1996
- PROVIDING OPTIMUM TORQUE TO IMPLANT PROSTHESESImplant Dentistry, 1993