Evaluating Rotational Stability of an Extended Depth of Focus Toric Intraocular Lens Using a Slit Lamp and Image-Based Analysis
Open Access
- 1 August 2020
- journal article
- research article
- Published by Taylor & Francis in Clinical Ophthalmology
- Vol. ume 14, 2405-2410
Abstract
Purpose: To evaluate the rotational stability of a toric extended depth of focus (EDOF) intraocular lens (IOL), using either slit lamp evaluation or image-processing software. Setting: Three clinical practices in the USA. Design: Prospective unmasked randomized clinical trial. Methods: Subjects presenting for routine cataract surgery that were interested in improved near vision received toric EDOF lenses (TECNIS Symfony® Toric) in both eyes. The measures of interest in the current analysis were the change in orientation of the IOL between 1 day, 1 month and 3 months postoperative. Orientation was measured at the microscope on the day of surgery, and with the slit lamp at all other visits. Day 1, 1-month and 3-month images of the lens orientation were captured with a slit lamp camera. Differences in orientation were recorded and analyzed. Results: A total of 150 eyes had IOL orientation data available. Image analysis showed mean absolute lens orientation changes from 1 day to 1 month and 3 months of less than 2 degrees. The percentage of lenses exhibiting rotation of ≤ 5 degrees between any measured time points was 97% or higher. Results were similar, but significantly more variable, when IOL orientation was measured at the slit lamp. Conclusion: The toric EDOF lens evaluated here demonstrated rotational stability that exceeded the prior ANSI standard. The best method to determine IOL orientation changes was through image analysis.Keywords
This publication has 14 references indexed in Scilit:
- Comparison of clinical outcomes of toric intraocular lens, Precizon vs Tecnis: a single center randomized controlled trialBMC Ophthalmology, 2018
- Rotational stability and visual performance 3 months after bilateral implantation of a new toric extended range of vision intraocular lensClinical Ophthalmology, 2018
- Comparison of Two Toric IOLs with Different Haptic Design: Optical Quality after 1 YearJournal of Ophthalmology, 2018
- Retreatments after multifocal intraocular lens implantation: an analysisClinical Ophthalmology, 2016
- Toric Intraocular Lenses in the Correction of Astigmatism During Cataract SurgeryOphthalmology, 2016
- Factors influencing 1-year rotational stability of AcrySof Toric intraocular lensesBritish Journal of Ophthalmology, 2015
- Clinical Outcomes of TECNIS Toric Intraocular Lens Implantation after Cataract Removal in Patients with Corneal AstigmatismOphthalmology, 2015
- Software-Based Evaluation of Toric IOL Orientation in a Multicenter Clinical StudyJournal of Refractive Surgery, 2014
- Long-term clinical outcomes of toric intraocular lens implantation in cataract cases with preexisting astigmatismJournal of Cataract & Refractive Surgery, 2014
- Calculating the optimal rotation of a misaligned toric intraocular lensJournal of Cataract & Refractive Surgery, 2008