Reply on RC1

Authors:

Abstract
The University of York’s laser-induced fluorescence (LIF) instrument for measuring sulfur dioxide (SO2) was compared to a commercial pulsed fluorescence (PF) and iodide chemical ionisation mass spectrometer (I-CIMS) aboard the UK FAAM research aircraft in both remote and ship-polluted marine environments. In high SO2 concentration plumes LIF and PF compared well, but LIF was the only instrument capable of SO2 measurements in the remote marine boundary layer due to its campaign limit of detection (LoD, 3 σ) of 0.07 ppb at 10 seconds compared with 0.4 ppb for the PF. Quantification of SO2 using I-CIMS was challenging due to a significant interference, but good signal correlation with the other instruments was observed in polluted air mases. A comparison of response time was also made, for which the I-CIMS and LIF proved much faster than the PF with 3-efolding times of 0.6, 2 and 17 seconds respectively. This work demonstrates the importance of sensitive instrumentation like the LIF for quantifying low concentrations of SO2, such as over remote marine environments, at the time resolutions required for a fast moving platform. This is particularly relevant now as a result of more stringent sulfur emission regulations for shipping, and likely more so in the future as anthropogenic SO2 concentrations continue to decline.

This publication has 41 references indexed in Scilit: