A Comparative Evaluation of Transcutaneous and End-Tidal Measurements of CO2 in Thoracic Anesthesia
Authors:
- Sungsam ChoSCSungsam Cho
- Tetsuya HaraTHTetsuya Hara
- Shiro TomiyasuSTShiro Tomiyasu
- Tetsuji MakitaTMTetsuji Makita
- Koji SumikawaKSKoji Sumikawa
- 1 September 2003
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 97 (3) , 776-779
Abstract
We performed this study to assess the accuracy of transcutaneous CO2 (PTCco2) monitoring compared with end-tidal CO2 (Petco2) in thoracic anesthesia. Twenty-six patients undergoing pneumonectomy with thoracotomy for which a long period of one-lung ventilation (OLV) was required were studied. The lungs were mechanically ventilated in the lateral decubitus position. PTCco2, Petco2, and arterial CO2 (Paco2) were simultaneously measured during two-lung ventilation (TLV) and during OLV at intervals of 15 min. All patients completed the study protocol. Bland-Altman analysis revealed a bias of −0.4 mm Hg with a precision of ±2.5 mm Hg during OLV and 1.4 mm Hg with ±4.3 mm Hg during TLV when PTCco2 and Paco2 were compared and revealed a bias of −5.8 mm Hg with a precision of ±4.1 mm Hg during OLV and −7.1 mm Hg with ±4.6 mm Hg during TLV when Petco2 and Paco2 were compared. We conclude that PTCco2 monitoring is accurate for evaluating CO2 levels during thoracic anesthesia.Keywords
This publication has 16 references indexed in Scilit:
- Noninvasive Monitoring of Carbon Dioxide During Mechanical Ventilation in Older Children: End-Tidal Versus Transcutaneous TechniquesAnesthesia & Analgesia, 2001
- Intraoperative Torsade de Pointes Ventricular Tachycardia and Ventricular Fibrillation During Sevoflurane AnesthesiaAnesthesia & Analgesia, 1998
- Variation in Arterial to End-Tidal CO2 Tension Differences During Anesthesia in the ???Kidney Rest??? Lateral Decubitus PositionAnesthesia & Analgesia, 1992
- A comparison of transcutaneous, end-tidal and arterial measurements of carbon dioxide during general anaesthesiaCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1992
- Accuracy of end-tidal PCO2 measurements using a sidestream capnometer in infants and children ventilated with the Sechrist infant ventilatorCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1990
- Carbon Dioxide MonitoringAnaesthesia and Intensive Care, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- DEADSPACE AND THE SINGLE BREATH TEST FOR CARBON DIOXIDE DURING ANAESTHESIA AND ARTIFICIAL VENTILATIONBritish Journal of Anaesthesia, 1984
- Cutaneous blood gas monitoring in the adultCritical Care Medicine, 1981
- Respiratory dead space and arterial to end-tidal CO2 tension difference in anesthetized manJournal of Applied Physiology, 1960