A Comparative Evaluation of Transcutaneous and End-Tidal Measurements of CO2 in Thoracic Anesthesia

Authors:
  • SCSungsam Cho
    Sungsam Cho
  • THTetsuya Hara
    Tetsuya Hara
  • STShiro Tomiyasu
    Shiro Tomiyasu
  • TMTetsuji Makita
    Tetsuji Makita
  • KSKoji Sumikawa
    Koji Sumikawa

Abstract
We performed this study to assess the accuracy of transcutaneous CO2 (PTCco2) monitoring compared with end-tidal CO2 (Petco2) in thoracic anesthesia. Twenty-six patients undergoing pneumonectomy with thoracotomy for which a long period of one-lung ventilation (OLV) was required were studied. The lungs were mechanically ventilated in the lateral decubitus position. PTCco2, Petco2, and arterial CO2 (Paco2) were simultaneously measured during two-lung ventilation (TLV) and during OLV at intervals of 15 min. All patients completed the study protocol. Bland-Altman analysis revealed a bias of −0.4 mm Hg with a precision of ±2.5 mm Hg during OLV and 1.4 mm Hg with ±4.3 mm Hg during TLV when PTCco2 and Paco2 were compared and revealed a bias of −5.8 mm Hg with a precision of ±4.1 mm Hg during OLV and −7.1 mm Hg with ±4.6 mm Hg during TLV when Petco2 and Paco2 were compared. We conclude that PTCco2 monitoring is accurate for evaluating CO2 levels during thoracic anesthesia.