Impact of reconstruction parameters on quantitative I-131 SPECT
- 28 June 2016
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 61 (14) , 5166-5182
Abstract
Radioiodine therapy using I-131 is widely used for treatment of thyroid disease or neuroendocrine tumors. Monitoring treatment by accurate dosimetry requires quantitative imaging. The high energy photons however render quantitative SPECT reconstruction challenging, potentially requiring accurate correction for scatter and collimator effects. The goal of this work is to assess the effectiveness of various correction methods on these effects using phantom studies. A SPECT/CT acquisition of the NEMA IEC body phantom was performed. Images were reconstructed using the following parameters: (1) without scatter correction, (2) with triple energy window (TEW) scatter correction and (3) with Monte Carlo-based scatter correction. For modelling the collimator-detector response (CDR), both (a) geometric Gaussian CDRs as well as (b) Monte Carlo simulated CDRs were compared. Quantitative accuracy, contrast to noise ratios and recovery coefficients were calculated, as well as the background variability and the residual count error in the lung insert. The Monte Carlo scatter corrected reconstruction method was shown to be intrinsically quantitative, requiring no experimentally acquired calibration factor. It resulted in a more accurate quantification of the background compartment activity density compared with TEW or no scatter correction. The quantification error relative to a dose calibrator derived measurement was found to be <1%,−26% and 33%, respectively. The adverse effects of partial volume were significantly smaller with the Monte Carlo simulated CDR correction compared with geometric Gaussian or no CDR modelling. Scatter correction showed a small effect on quantification of small volumes. When using a weighting factor, TEW correction was comparable to Monte Carlo reconstruction in all measured parameters, although this approach is clinically impractical since this factor may be patient dependent. Monte Carlo based scatter correction including accurately simulated CDR modelling is the most robust and reliable method to reconstruct accurate quantitative iodine-131 SPECT images.Keywords
This publication has 33 references indexed in Scilit:
- A Parallel-Cone Collimator for High-Energy SPECTJournal of Nuclear Medicine, 2015
- Improving quantitative dosimetry in 177Lu-DOTATATE SPECT by energy window-based scatter correctionsNuclear Medicine Communications, 2014
- MIRD Pamphlet No. 24: Guidelines for Quantitative 131I SPECT in Dosimetry ApplicationsJournal of Nuclear Medicine, 2013
- Quantitative SPECT/CT: SPECT joins PET as a quantitative imaging modalityEuropean Journal of Nuclear Medicine and Molecular Imaging, 2013
- Guidelines for Imaging and Staging of Neuroblastic Tumors: Consensus Report from the International Neuroblastoma Risk Group ProjectRadiology, 2011
- Hyperthyroidism and Other Causes of Thyrotoxicosis: Management Guidelines of the American Thyroid Association and American Association of Clinical EndocrinologistsThyroid®, 2011
- Monte Carlo-based statistical SPECT reconstruction: influence of number of photon tracksIEEE Transactions on Nuclear Science, 2005
- Efficient fully 3-D iterative SPECT reconstruction with Monte Carlo-based scatter compensationIEEE Transactions on Medical Imaging, 2002
- Acceleration of Monte Carlo SPECT simulation using convolution-based forced detectionIEEE Transactions on Nuclear Science, 2001
- Quantitative /sup 131/I SPECT with triple energy window Compton scatter correctionIEEE Transactions on Nuclear Science, 1998