Temporal changes in sucrose phantoms for magnetic resonance imaging: effects of preparation temperatures
Authors:
- 29 July 2025
- journal article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 18 (4) , 1283-1293
Abstract
This study investigated the effects of preparation temperature and usage period on the relaxation times and apparent diffusion coefficients (ADC) of sucrose phantoms to enhance imaging reliability. Phantoms were prepared using 10% sucrose solutions at 20, 50, and 80 °C. T1 and T2 relaxation times and ADC were monitored over 80 days using magnetic resonance imaging on a 1.5 T system. T2 relaxation time in 50 °C solutions increased from 245.5 to 1579 ms, while 80 °C solutions showed the highest stability (coefficient of variation ≈ 1.8%). T1 relaxation time changes were minimal, and ADC decreased at an average rate of 2.19 × 10-6 mm2/s per day. Bacteria were observed in the sucrose solution, and higher protein concentrations were strongly correlated with decreased 1/T2. In conclusion, sucrose phantoms exhibited temperature-dependent stability, with 80 °C preparations providing the most reliable T2 relaxation time.Keywords
This publication has 30 references indexed in Scilit:
- Feasibility of accelerated simultaneous multislice diffusion‐weighted MRI of the prostateJournal of Magnetic Resonance Imaging, 2017
- A phantom for diffusion-weighted MRI (DW-MRI)Journal of Magnetic Resonance Imaging, 2013
- Multi‐system repeatability and reproducibility of apparent diffusion coefficient measurement using an ice‐water phantomJournal of Magnetic Resonance Imaging, 2012
- Measurement of longitudinal (T1) relaxation in the human lung at 3.0 Tesla with tissue‐based and regional gradient analysesJournal of Magnetic Resonance Imaging, 2007
- Quantitative apparent diffusion coefficients and T2 relaxation times in characterizing contrast enhancing brain tumors and regions of peritumoral edemaJournal of Magnetic Resonance Imaging, 2005
- Sucrose utilisation in bacteria: genetic organisation and regulationApplied Microbiology and Biotechnology, 2005
- Modified Look‐Locker inversion recovery (MOLLI) for high‐resolution T1 mapping of the heartMagnetic Resonance in Medicine, 2004
- Tissue‐equivalent phantoms for high frequenciesConcepts in Magnetic Resonance Part B, 2004
- Fast T1 mapping on a whole-body scannerMagnetic Resonance in Medicine, 1999
- A Phantom for diffusion‐weighted imaging of acute strokeJournal of Magnetic Resonance Imaging, 1998