Comparison of Predictability in Vault Using NK Formula and KS Formula for the Implantable Collamer Lens Surgery
Open Access
- 1 January 2024
- journal article
- research article
- Published by Wiley in Journal of Ophthalmology
- Vol. 2024 (1) , 4256371
Abstract
Background. This study aims to investigate the agreement between the NK and KS formulas in predicting the vault after implantation of an EVO‐implantable collamer lens (ICL). Methods. This retrospective study included 106 eyes of 57 patients who underwent ICL‐V4c implantation. Preoperative vault prediction was conducted by utilizing the NK and KS formulas, with postoperative measurements by anterior segment optical coherence tomography (AS‐OCT) at one month. The analysis focused on the consistency between predicted and achieved vaults, as well as the correlation between the achieved vault and various biometric parameters. Results. The mean achieved vault was 605.25 ± 212.72 µm, which was significantly smaller than the predicted vaults of 710.08 ± 195.08 and 673.80 ± 212.76 µm, using the NK and KS formulas, respectively (P < 0.05). The mean differences between the achieved vault and the predicted vault using the NK formula and KS formula were −104.82 μm (95% LoA: −600.38–391.19 μm) and −68.55 μm (95% LoA: −628.91–491.82 μm), respectively. Anterior chamber depth (ACD), vertical sulcus‐to‐sulcus (V‐STS) diameter, and crystalline lens rise (CLR) were independent factors associated with the achieved vault (P < 0.05). The two formulas showed no statistically significant difference in absolute prediction error (APE). Conclusion. The NK formula exhibited superior consistency and low predictive error compared to the KS formula in the 12.6 mm ICL group. AS‐OCT measurements overestimated the predicted ICL vault, especially in the 13.2 mm ICL size selection. Relying solely on the NK or KS formulas for predicting vaults before ICL surgery is not recommended.Keywords
All Related Versions
Funding Information
- Guangdong Medical Research Foundation (B2023206)
- Shantou Science and Technology Project (190917155269927)
- Li Ka Shing Foundation (2020LKSFG06B)
This publication has 32 references indexed in Scilit:
- Predictability of the vault after posterior chamber phakic intraocular lens implantation using anterior segment optical coherence tomographyJournal of Cataract & Refractive Surgery, 2019
- Implantable Collamer Lens Sizing Method Based on Swept-Source Anterior Segment Optical Coherence TomographyAmerican Journal of Ophthalmology, 2018
- Clinical Outcomes and Cataract Formation Rates in Eyes 10 Years After Posterior Phakic Lens Implantation for MyopiaJAMA Ophthalmology, 2016
- Optimization of an Implantable Collamer Lens Sizing Method Using High-Frequency Ultrasound BiomicroscopyAmerican Journal of Ophthalmology, 2012
- Correlation between Preoperative Biometry and Posterior Chamber Phakic Visian Implantable Collamer Lens VaultingOphthalmology, 2012
- Improving accuracy of phakic intraocular lens sizing using high-frequency ultrasound biomicroscopyJournal of Cataract & Refractive Surgery, 2011
- Direct Measurement of the Ciliary Sulcus Diameter by 35-Megahertz Ultrasound BiomicroscopyOphthalmology, 2007
- Anterior chamber width measurement by high-speed optical coherence tomographyOphthalmology, 2005
- United States Food and Drug Administration clinical trial of the Implantable Collamer Lens (ICL) for moderate to high myopiaOphthalmology, 2004
- Predicting sulcus size using ocular measurementsJournal of Cataract & Refractive Surgery, 2001