Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System
Open Access
- 2 February 2018
- journal article
- Published by Termedia Sp. z.o.o. in Polish Journal of Radiology
- Vol. 82, 498-505
Abstract
Background: Cardiac magnetic resonance imaging (cMRI) has become the non-invasive reference standard for the evaluation of cardiac function and viability. The introduction of open, high-field, 1.0T (HFO) MR scanners offers advantages for examinations of obese, claustrophobic and paediatric patients. The aim of our study was to compare standard cMRI sequences from an HFO scanner and those from a cylindrical, 1.5T MR system. Material/Method: Fifteen volunteers underwent cMRI both in an open HFO and in a cylindrical MR system. The protocol consisted of cine and unenhanced tissue sequences. The signal-to-noise ratio (SNR) for each sequence and blood-myocardium contrast for the cine sequences were assessed. Image quality and artefacts were rated. The location and number of non-diagnostic segments was determined. Volunteers’ tolerance to examinations in both scanners was investigated. Results: SNR was significantly lower in the HFO scanner (all p<0.001). However, the contrast of the cine sequence was significantly higher in the HFO platform compared to the 1.5T MR scanner (0.685±0.41 vs. 0.611±0.54; p0.05). Overall, only few non-diagnostic myocardial segments were recorded: 6/960 (0.6%) by the HFO and 17/960 (1.8%) segments by the cylindrical system. The volunteers expressed a preference for the open MR system (p<0.01). Conclusions: Standard cardiac MRI sequences in an HFO platform offer a high image quality that is comparable to the quality of images acquired in a cylindrical 1.5T MR scanner. An open scanner design may potentially improve tolerance of cardiac MRI and therefore allow to examine an even broader patient spectrum.Keywords
This publication has 25 references indexed in Scilit:
- Feasibility of High-Dose Dobutamine Stress SSFP Cine MRI at 3 Tesla with Patient Adaptive Local RF Shimming using Dual-Source RF Transmission: Initial ResultsRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2012
- Bright-Blood T2-Weighted MRI Has Higher Diagnostic Accuracy Than Dark-Blood Short Tau Inversion Recovery MRI for Detection of Acute Myocardial Infarction and for Assessment of the Ischemic Area at Risk and Myocardial SalvageCirculation: Cardiovascular Imaging, 2011
- ACCF/ACR/AHA/NASCI/SCMR 2010 Expert Consensus Document on Cardiovascular Magnetic ResonanceJACC, 2010
- Empfehlungen für den Einsatz der Computertomografie und Magnetresonanztomografie in der Herzdiagnostik. Teil 2 – MagnetresonanztomografieRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2009
- Cardiac Magnetic Resonance Imaging Using an Open 0.35 T SystemJournal of Computer Assisted Tomography, 2007
- Improved turbo spin‐echo imaging of the heart with motion‐trackingJournal of Magnetic Resonance Imaging, 2006
- Comprehensive Cardiac Magnetic Resonance Imaging at 3.0 TeslaInvestigative Radiology, 2006
- Cardiac Steady-State Free Precession CINE Magnetic Resonance Imaging at 3.0 TeslaInvestigative Radiology, 2006
- In vivo multiple-mouse imaging at 1.5 TMagnetic Resonance in Medicine, 2003
- k‐Space interpretation of the Rose Model: Noise limitation on the detectable resolution in MRIMagnetic Resonance in Medicine, 2002