Towards a quantum cascade laser-based implant for the continuous monitoring of glucose
- 17 October 2018
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in The Analyst
- Vol. 143 (24) , 6025-6036
Abstract
Continuous glucose monitoring enables an improved disease management for people with diabetes. However, state-of-the-art, enzyme-based, minimally invasive sensors lose their sensitivity over time and have to be replaced periodically. Here, we present the in vitro investigation of a quantum cascade laser-based measurement scheme that conceptually should be applicable over elongated periods of time due to its reagent-free nature and may therefore be considered as an approach towards long-term implantation. The method uses a miniaturized optofluidic interface in transflection geometry to measure the characteristic mid-infrared absorption properties of glucose. A glucose sensitivity of 3.2 mg dL−1 is achieved in aqueous glucose solutions. While this sensitivity drops to 12 mg dL−1 in the presence of biologically plausible, maximum concentrations of other monosaccharides, it is still well within the medically acceptable range according to Parkes error grid analysis. With a response time of less than five minutes, our sensor should be able to react adequately fast to physiological changes in glucose concentration. Finally, no drift or deterioration was found during an extended, 42 days in vitro experiment. These results underline the potential of this technique for its conceivable applicability in vivo as a long-term glucose monitoring implant.Keywords
Funding Information
- Heidelberg Graduate School of Fundamental Physics
This publication has 34 references indexed in Scilit:
- In vitro measurements of physiological glucose concentrations in biological fluids using mid-infrared lightBiomedical Optics Express, 2013
- The inherent problem of transflection-mode infrared spectroscopic microscopy and the ramifications for biomedical single point and imaging applicationsThe Analyst, 2012
- Continuous glucose monitoring by means of mid-infrared transmission laser spectroscopy in vitroThe Analyst, 2011
- Tunable external cavity quantum cascade laser for the simultaneous determination of glucose and lactate in aqueous phaseThe Analyst, 2010
- Theory of Midinfrared Absorption Microspectroscopy: I. Homogeneous SamplesAnalytical Chemistry, 2010
- A Study of Electric Field Standing Waves on Reflection Microspectroscopy of Polystyrene ParticlesApplied Spectroscopy, 2009
- Middle Infrared, Quantum Cascade Laser Optoelectronic Absorption System for Monitoring Glucose in SerumApplied Spectroscopy, 2005
- Towards functional group-specific detection in high-performance liquid chromatography using mid-infrared quantum cascade lasersJournal of Chromatography A, 2001
- An Introduction to Multivariate Calibration and AnalysisAnalytical Chemistry, 1987
- Optical constants of water in the infrared: Influence of temperature*Journal of the Optical Society of America, 1977