Effectiveness of Erythrocyte Morphology Observation as an Indicator for the Selection and Qualification of Blood in a Mechanically Induced Hemolysis Test
Open Access
- 29 May 2024
- journal article
- research article
- Published by MDPI AG in Applied Sciences
- Vol. 14 (11) , 4695
Abstract
Background: This study was conducted to confirm the reliability of an in vitro mechanically induced hemolysis test (ISO 10993-4:2017), which is essential for ensuring the safety of blood pumps. Methods: For appropriate anticoagulant selection, porcine blood was prepared in anticoagulant citrate dextrose solution A (ACD-A), heparin, and citrate phosphate dextrose adenine (CPDA-1), respectively, according to the ASTM F1830 standard. Anticoagulant-treated porcine and bovine blood were circulated in a mock circulatory loop (MCL) for 6 h to observe the rate of plasma-free hemoglobin (pfHb) and RBCs with morphological integrity. Results: A morphological loss of red blood cells (RBCs) was observed over time. While there were differences in morphological loss depending on the anticoagulant, no consistent trend could be identified. The pfHb concentration was significantly higher in bovine than in porcine blood. Conversely, the number of RBCs with morphological integrity decreased over time in both, but the ratio of RBCs with morphological integrity was similar across all timepoints. Conclusions: The percentage of RBCs with morphological integrity can be used as a reliable indicator for the interpretation of mechanically induced hemolysis results in different blood types. Furthermore, the reliability of the in vitro mechanically induced hemolysis test (ISO 10993-4:2017) was assessed.Keywords
Funding Information
- Ministry of Food and Drug Safety (21174MFDS242)
This publication has 24 references indexed in Scilit:
- Potential Factors for Poor Reproducibility of In Vitro Hemolysis TestingASAIO Journal, 2021
- Validation of a Miniaturized Test Loop for the Assessment of Human Blood Damage by Continuous-Flow Left-Ventricular Assist DevicesAnnals of Biomedical Engineering, 2021
- Miniaturized Test Loop for the Assessment of Blood Damage by Continuous-Flow Left-Ventricular Assist DevicesAnnals of Biomedical Engineering, 2019
- The CentriMag Centrifugal Blood Pump as a Benchmark for In Vitro Testing of Hemocompatibility in Implantable Ventricular Assist DevicesArtificial Organs, 2014
- Biomaterial testing for covered stent membranes: Evaluating thrombosis and restenosis potentialJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2011
- Hemolysis During Cardiac Extracorporeal Membrane Oxygenation: A Case-Control Comparison of Roller Pumps and Centrifugal Pumps in a Pediatric PopulationASAIO Journal, 2011
- Comparison of the centrifugal and roller pump in elective coronary artery bypass surgery—a prospective, randomized study with special emphasis upon platelet activationScandinavian Cardiovascular Journal, 2003
- Superiority of centrifugal pump over roller pump in paediatric cardiac surgery: prospective randomised trialEuropean Journal of Cardio-Thoracic Surgery, 1998
- Centrifugal pump and heparin coating improves cardiopulmonary bypass biocompatibilityThe Annals of Thoracic Surgery, 1996
- Relationship of hyperthermia-induced hemolysis of human erythrocytes to the thermal denaturation of membrane proteinsBiochimica et Biophysica Acta (BBA) - Biomembranes, 1989