MTF measurement and a phantom study for scatter correction in CBCT using primary modulation
Authors:
- 1 January 2006
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- Vol. 3, 1964-1968
Abstract
Recently, we proposed a scatter correction methods for X-ray imaging using primary modulation. A primary modulator with spatially variant attenuating materials is inserted between the X-ray source and the object to make the scatter and part of the primary distributions strongly separated in the Fourier domain. Linear filtering and demodulation techniques suffice to extract and correct the scatter for this modified system. The method has been verified by computer simulations and preliminary experimental results. In this work, we look into a hybrid method using both primary modulation and an anti-scatter grid. The reconstructed image resolution using the proposed approach is evaluated by MTF measurements, and the scatter correction performance is also investigated by experiments on a human chest phantom. The results using the proposed hybrid method are compared with those using an antiscatter grid only. Experiments with scatter inherently suppressed using a narrowly opened collimator and an anti-scatter grid (a slot-scan geometry) were also carried out. The comparison shows that the filtering in the proposed algorithm does not impair the image resolution, and the primary modulation method can effectively suppress the scatter artifacts. In the central region of interest, the reconstruction error relative to the image obtained using a slot-scan geometry is reduced from 18.10% to 0.48% for the human chest phantom, if the primary modulation method is used.Keywords
This publication has 18 references indexed in Scilit:
- Estimating 0thand 1stmoments in C-arm CT data for extrapolating truncated projections2005
- X-ray scatter correction for cone-beam CT using moving blocker array2005
- X‐ray scatter correction algorithm for cone beam CT imagingMedical Physics, 2004
- High resolution anthropomorphic phantom for Monte Carlo analysis of internal radiation sources2002
- X-ray scatter suppression algorithm for cone-beam volume CT2002
- Scatter compensation in digital chest radiography using the posterior beam stop techniqueMedical Physics, 1994
- Scatter Compensation for Digital Chest Radiography Using Maximum Likelihood Expectation MaximizationInvestigative Radiology, 1993
- X‐ray scatter removal by deconvolutionMedical Physics, 1988
- Scatter estimation for a digital radiographic system using convolution filteringMedical Physics, 1987
- Practical cone-beam algorithmJournal of the Optical Society of America A, 1984