Abrasive Properties of Braided Polyblend Sutures in Cuff Tendon Repair: An in Vitro Biomechanical Study Exploring Regular and Tape Sutures
Authors:
- 21 August 2014
- journal article
- Published by Wiley in Arthroscopy
- Vol. 30 (12) , 1569-1573
Abstract
Purpose: This study aimed to evaluate the abrasive properties of different suture materials (tape or regular) on the infraspinatus tendon of sheep. Methods: Four types of sutures were compared: FiberWire (Arthex, Naples FL), FiberTape (Arthrex), Orthocord (DePuy Mitek, Raynham, MA) and ForceFiber (Tornier, Bloomington, MN). Each suture (n = 10) was cycled with a filxed load of 10 N and an alternating motion of the suture through sheep infraspinatus tendon, with an excursion of 30 mm. The migration of the suture as it cut through the tissue was measured at intervals of 5 cycles, up to failure or a total of 50 cycles, or a tendon tear greater than 13 mm. Results: ForceFiber and Orthocord sutures showed a significantly ( P < .05) lower amount of abrasion compared with FiberWire and FiberTape: The mean cutting rate (defined as the size of the defect at the end of the test divided by 50 when this number of cycles was reached, or as 13 mm divided by the number of cycles to reach this value when the test was stopped before 50 cycles) was, respectively, 0.04 mm/cycle, 0.12 mm/cycle, 0.11 mm/cycle, 0.32 mm/cycle, and 0.25 mm/cycle. The defect size at 15 cycles was, respectively, 5.7 mm, 5.6 mm, 9.4 mm, 7.7 mm, and 7.4 mm. Although no statistical significance was found, sutures shaped in a tape form (FiberTape) were less aggressive on the tendon than the corresponding sutures in regular form (FiberWire). Conclusions: This study found increased abrasive effects of FiberWire and FiberTape compared with ForceFiber and Orthocord sutures. Clinical Relevance: Currently, surgeons have a large choice of suture materials. Knowledge of biomechanical characteristics of different braided polyblend suture materials could help surgeons decide which suture to use for rotator cuff tears.Keywords
This publication has 12 references indexed in Scilit:
- Influence of prolonged immersion on the resistance of arthroscopy knots in biological mediaOrthopaedics & Traumatology: Surgery & Research, 2013
- Tendon fixation in arthroscopic latissimus dorsi transfer for irreparable posterosuperior cuff tears: An in vitro biomechanical comparison of interference screw and suture anchorsClinical Biomechanics, 2011
- Animal Models for Translational Research on Shoulder PathologiesSports Medicine and Arthroscopy Review, 2011
- Influence of Suture Material on the Biomechanical Behavior of Suture-Tendon SpecimensThe American Journal of Sports Medicine, 2008
- Evaluation of Suture Abrasion Against Rotator Cuff Tendon and Proximal Humerus BoneArthroscopy, 2007
- Glove Tears During Arthroscopic Shoulder Surgery Using Solid‐Core SutureArthroscopy, 2007
- Mechanical and Handling Properties of Braided Polyblend Polyethylene Sutures in Comparison to Braided Polyester and Monofilament Polydioxanone SuturesArthroscopy, 2006
- Abrasion resistance of two types of nonabsorbable braided sutureArthroscopy, 2004
- Failure of suture material at suture anchor eyeletsArthroscopy, 2002
- Full-Thickness Rotator Cuff TearsThe American Journal of Sports Medicine, 1996