Epiretinale Sehprothesen
Authors:
- 1 October 2005
- journal article
- research article
- Published by Springer Nature in Der Ophthalmologe
- Vol. 102 (10) , 933-940
Abstract
Epiretinal implants consist of a camera chip capturing the scene, a visual processor calculating retina-specific pulse sequences, a transponder for data and energy, and the implant itself. The implant consists of a receiver integrated into a posterior chamber lens, a microcable, and the retina stimulator. The stimulator is fixated onto the retinal surface using retinal tacks. In animal experiments it was shown that the materials being used were tolerated, the data and energy stream did not induce any adverse events, and that the fixation was feasible using retinal tacks. Stimulation of the inner retinal surface yielded a topographically correct activation of visual cortical areas. Even after a long history of blindness, tests in humans disclosed visual percepts.Keywords
This publication has 32 references indexed in Scilit:
- A wavelet-based time–frequency analysis approach for classification of motor imagery for brain–computer interface applicationsJournal of Neural Engineering, 2005
- Cortical Activation Via an Implanted Wireless Retinal ProsthesisInvestigative Ophthalmology & Visual Science, 2005
- Long-term stimulation by active epiretinal implants in normal and RCD1 dogsJournal of Neural Engineering, 2005
- MEMS technologies for epiretinal stimulation of the retinaJournal of Micromechanics and Microengineering, 2004
- The Retina Implant New Approach to a Visual Prosthesis / Das Retina-Implant Ein neuer Ansatz für eine SehprotheseBiomedical Engineering / Biomedizinische Technik, 2004
- Sputtered Iridium Oxide Films as Charge Injection Material for Functional ElectrostimulationJournal of the Electrochemical Society, 2004
- Micro-transponder systems for medical applicationsIEEE Transactions on Instrumentation and Measurement, 2001
- Retinal prosthesisOphthalmology, 2001
- SUCCESSFUL LONG-TERM IMPLANTATION OF ELECTRICALLY INACTIVE EPIRETINAL MICROELECTRODE ARRAYS IN RABBITSRetina, 1999
- SUCCESSFUL LONG-TERM IMPLANTATION OF ELECTRICALLY INACTIVE EPIRETINAL MICROELECTRODE ARRAYS IN RABBITSRetina, 1999