Energy response of EBT3 radiochromic films: implications for dosimetry in kilovoltage range
Open Access
- 6 January 2014
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 15 (1) , 331-338
Abstract
The objective of this study is to evaluate the suitability of recently introduced radiochromic film EBT3 for clinical dosimetry in the kilovoltage (kV) range. For this purpose, a kV X-ray irradiator, X RAD 320ix in the range 70 to 300 kVp, a clinical source, a 6 MV and an 18 MV X-ray clinical beam from a Varian linear accelerator were calibrated following AAPM dosimetry protocols. EBT3 films from two different EBT3 batches were placed side-by-side on the surface of a water phantom; doses from 0.5 to 4 Gy were delivered. Similarly, irradiations were performed for and 6 and 18 MV beams in a water-equivalent phantom. Films were reproducibly placed at the center of a flatbed scanner and 48-bit RGB scans were obtained both pre- and postirradiations. Net optical density (netOD) and response for a given radiation quality relative to was determined for each EBT3 film. The netOD of the red color showed reproducible response (within 1%) for both batches when irradiated using the source. For a given dose of 1 Gy of kVp X-ray, the response relative to using the three color channels (red, green, and blue) decreases with decrease in kVp, reaching a maximum underresponse of for the 70 kVp. A significant underresponse of was observed at 300 kVp. Responses of MV X-ray beams with respect to at the 1 Gy dose level showed no statistically significant difference. A relatively small difference in the response was observed between the two EBT3 batches used in this study in the kV X-ray range. PACS numbers: 87.56.B, 87.57.uq
Keywords
This publication has 9 references indexed in Scilit:
- Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verificationJournal of Applied Clinical Medical Physics, 2013
- Dose‐response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron‐produced monochromatic x‐ray beamsMedical Physics, 2012
- Linearization of dose–response curve of the radiochromic film dosimetry systemMedical Physics, 2012
- Multichannel film dosimetry with nonuniformity correctionMedical Physics, 2011
- Monte Carlo calculated absorbed‐dose energy dependence of EBT and EBT2 filmMedical Physics, 2010
- Investigation of energy dependence of EBT and EBT‐2 Gafchromic filmMedical Physics, 2010
- Precise radiochromic film dosimetry using a flat-bed document scannerMedical Physics, 2005
- AAPM protocol for 40–300 kV x‐ray beam dosimetry in radiotherapy and radiobiologyMedical Physics, 2001
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999