Abstract
The room-temperature shimming coils are installed within the bore of the superconducting magnet. By applying different excitation currents to each set of coils, different compensating magnetic fields are generated to modify specific components of the main magnetic field, thereby improving its uniformity. For a nuclear magnetic resonance spectrometer operating at 400 MHz, the initial magnetic field uniformity must be better than 1 ppm in a 10 mm DSV (Diameter Spherical Volume). After active shimming using the room-temperature shimming coils, the uniformity should be improved to within 10 ppb in the same DSV. This paper focuses on the design of a room-temperature shimming system for a 400 MHz NMR spectrometer. The coil design for each shimming order is based on spherical harmonic function analysis and the target field method, implemented via the finite difference stream function approach. Additionally, the measurement system and power supply system for the shimming coils are described.