Dynamic Biomechanical Examination of the Lumbar Spine With Implanted Total Disc Replacement Using a Pendulum Testing System
Authors:
- Alan H. DanielsADAlan H. Daniels
- David J. PallerDPDavid J. Paller
- Sarath KoruproluSKSarath Koruprolu
- Matthew McDonnellMMMatthew McDonnell
- Mark A. PalumboMPMark A. Palumbo
- Joseph J. CriscoJCJoseph J. Crisco
- 1 November 2012
- journal article
- research article
- Published by Wolters Kluwer Health in Spine
- Vol. 37 (23) , E1438-E1443
Abstract
Study Design. Biomechanical cadaver investigation. Objective. To examine dynamic bending stiffness and energy absorption of the lumbar spine with and without implanted total disc replacement (TDR) under simulated physiological motion. Summary of Background Data. The pendulum testing system is capable of applying physiological compressive loads without constraining motion of functional spinal units (FSUs). The number of cycles to equilibrium observed under pendulum testing is a measure of the energy absorbed by the FSU. Methods. Five unembalmed, frozen human lumbar FSUs were tested on the pendulum system with axial compressive loads of 181 N, 282 N, 385 N, and 488 N before and after Synthes ProDisc-L TDR implantation. Testing in flexion, extension, and lateral bending began by rotating the pendulum to 5º resulting in unconstrained oscillatory motion. The number of rotations to equilibrium was recorded and bending stiffness (N·m/º) was calculated and compared for each testing mode. Results. In flexion/extension, the TDR constructs reached equilibrium with significantly (P < 0.05) fewer cycles than the intact FSU with compressive loads of 282 N, 385 N, and 488 N. Mean dynamic bending stiffness in flexion, extension, and lateral bending increased significantly with increasing load for both the intact FSU and TDR constructs (P < 0.001). In flexion, with increasing compressive loading from 181 N to 488 N, the bending stiffness of the intact FSUs increased from 4.0 N·m/º to 5.5 N·m/º, compared with 2.1 N·m/º to 3.6 N·m/º after TDR implantation. At each compressive load, the intact FSU was significantly stiffer than the TDR (P < 0.05). Conclusion. Lumbar FSUs with implanted TDR were found to be less stiff, but absorbed more energy during cyclic loading with an unconstrained pendulum system. Although the effects on clinical performance of motion-preserving devices are not fully known, these results provide further insight into the biomechanical behavior of these devices under approximated physiological loading conditions.Keywords
This publication has 36 references indexed in Scilit:
- A 6.5-year follow-up of 14 patients who underwent ProDisc total disc arthroplasty for combined long-standing degenerative lumbar disc disease and recent disc herniationJournal of Clinical Neuroscience, 2011
- Optimal stiffness of a pedicle-screw-based motion preservation implant for the lumbar spineEuropean Spine Journal, 2011
- Does anterior lumbar interbody fusion promote adjacent degeneration in degenerative disc disease? A finite element studyJournal of Orthopaedic Science, 2011
- Effect of multilevel lumbar disc arthroplasty on spine kinematics and facet joint loads in flexion and extension: a finite element analysisEuropean Spine Journal, 2010
- In Situ Contact Analysis of the Prosthesis Components of Prodisc-L in Lumbar Spine Following Total Disc ReplacementSpine, 2009
- Biomechanical Effect of Constraint in Lumbar Total Disc ReplacementSpine, 2009
- Effect of an artificial disc on lumbar spine biomechanics: a probabilistic finite element studyEuropean Spine Journal, 2008
- Early stage disc degeneration does not have an appreciable affect on stiffness and load transfer following vertebroplasty and kyphoplastyEuropean Spine Journal, 2008
- Dynamics of an Intervertebral Disc Prosthesis in Human Cadaveric SpinesHSS Journal®: The Musculoskeletal Journal of Hospital for Special Surgery, 2007
- Investigation of Vibration Characteristics of the Ligamentous Lumbar Spine Using the Finite Element ApproachJournal of Biomechanical Engineering, 1994