MR-guided Radiofrequency Ablation: Do Magnetic Fields Influence Extent of Coagulation in ex Vivo Bovine Livers?

Authors:
  • CAChristophe Aubé
    Christophe Aubé
  • HGHansjoerg Graf
    Hansjoerg Graf
  • ABAndreas Boss
    Andreas Boss
  • SCStephan Clasen
    Stephan Clasen
  • THTina Herberts
    Tina Herberts
  • DSDiethard Schmidt
    Diethard Schmidt
  • FSFritz Schick
    Fritz Schick
  • CCClaus D. Claussen
    Claus D. Claussen

Abstract
Purpose: To prospectively determine if static magnetic fields of magnetic resonance (MR) imagers affect radiofrequency (RF) ablation coagulation volume and shape. Materials and Methods: Ex vivo RF ablations of bovine livers were performed with magnetic field strengths of 0.2, 1.5, and 3.0 T and were compared with ablations performed outside the magnetic field in a control group. Two MR-compatible monopolar RF devices (internally cooled single and cluster electrodes) were systematically tested. Length of long axis (y-axis), length of two short axes (x- and z-axes), and coagulation volume and shape measured outside and inside different magnetic fields were compared with the Dunnett test. Significance level was set to .05. Results: For the single electrode, no significant difference was observed between length of short axes and coagulation volume and shape measured inside and outside the magnetic field. Mean x- and z-axis lengths were 2.3 and 2.6 cm, respectively, outside the magnetic field; 2.4 and 2.4 cm, ...