Performance Evaluation of Platelet Counting by Novel Fluorescent Dye Staining in the XN-Series Automated Hematology Analyzers
- 19 March 2014
- journal article
- Published by Wiley in Journal of Clinical Laboratory Analysis
- Vol. 28 (5) , 341-348
Abstract
Conventional automated hematology analyzers have limitations in platelet measurements such as poor accuracy and precision in the low count range and interference by nonplatelet particles. In order to improve it, the newly developed XN-Series automated hematology analyzers (Sysmex Corporation, Kobe, Japan) have been installed with a new dedicated channel for platelet analysis (PLT-F), which is based on a fluorescence flow cytometry method with uses of a novel fluorescent dye specifically staining platelets. We evaluated the basic performance of this new PLT-F channel. Basic performance of the PLT-F channel in within-run reproducibility and assay linearity was studied using standard methods. Correlation was studied between PLT-F and a conventional automated hematology analyzer (XE-2100) and immunoplatelet analysis using anti-CD61 monoclonal antibody (Cell-Dyn Sapphire; Abbott Laboratories). The assay interference by nonplatelet particles such as fragmented red and white blood cells was evaluated by using clinical samples, respectively, from burn injury and acute leukemia. Basic performance of the PLT-F platelet counting was satisfactory in within-run reproducibility, linearity and correlation with the conventional analyzer. The correlation was satisfactory also with the immunoplatelet analysis, even for samples from a patient with burn injury, and those with white blood cell fragments displayed, platelet abnormal flag and low platelet counts (<50 × 10(9)/l). The platelet counting performance of the PLT-F channel of the XN Series had improved accuracy and precision in the low range and in abnormal samples, avoiding the interference by nonplatelet particles.Keywords
This publication has 11 references indexed in Scilit:
- Performance evaluation of the Sysmex haematology XN modular systemJournal of Clinical Pathology, 2012
- Evaluation of the platelet counting by Abbott CELL-DYN®SAPPHIRE™ haematology analyser compared with flow cytometryInternational Journal of Laboratory Hematology, 2009
- Leukemia burden delays lymphocyte and platelet recovery after allo-SCT for AMLBone Marrow Transplantation, 2008
- Performance of automated platelet quantification by using evoluted analyser in comparison with a immunological reference methods in thrombocytopenic patients2008
- Accuracy of platelet counting haematology analysers in severe thrombocytopenia and potential impact on platelet transfusionBritish Journal of Haematology, 2005
- The most accurate platelet count on the Sysmex XE-2100. Optical or impedance?Clinical and Laboratory Haematology, 2004
- An Interlaboratory Study of a Candidate Reference Method for Platelet CountingAmerican Journal of Clinical Pathology, 2001
- New quantitative parameters on a recently introduced automated blood cell counter - the XE 2100TMClinical and Laboratory Haematology, 2000
- A Rapid and Accurate Closed-Tube Immunoassay for Platelets on an Automated Hematology AnalyzerAmerican Journal of Clinical Pathology, 2000
- Spurious elevation of automated platelet counts in secondary acute monocytic leukemia associated with tumor lysis syndrome.1999