Characterization of Nova plasmas using an x-ray spectrometer with temporal and spatial resolution
Authors:
- 1 January 1995
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 66 (1) , 764-766
Abstract
Spectroscopic diagnostics have great potential to obtain high temperature measurements of plasmas created in inertial confinement fusion targets. The plasmas may be over 1 mm in size and therefore, one of the first steps in making better spectroscopic measurements has been to develop a spectrograph which takes advantage of gated technology by coupling a Bragg crystal to a microchannel plate. The crystal disperses the x rays, while slits add the ability to image the plasmas in the perpendicular direction. The characteristics of this diagnostic, TSPEC, are evaluated for laser‐produced plasmas. Recent data will be presented from colliding plasmas and large‐scale hohlraums which indicate that imaging can greatly enhance the ability to diagnose these plasmas.Keywords
This publication has 14 references indexed in Scilit:
- A new multichannel soft x-ray framing camera for fusion experimentsReview of Scientific Instruments, 1992
- Electron-temperature measurement in laser-produced plasmas by the ratio of isoelectronic line intensitiesPhysical Review A, 1992
- Characterization of a laser-produced plasma using the technique of point-projection absorption spectroscopyPhysical Review A, 1991
- High speed proximity focused X-ray camerasLaser and Particle Beams, 1991
- Kαabsorption spectroscopy: Diagnostic of the radiative preheating of a laser-irradiated layered targetPhysical Review A, 1989
- X-ray absorption spectroscopy of laser-produced plasmas: A study of the experiment and data analysisPhysical Review A, 1989
- Time-Resolved Measurement of Coronal Temperature and Line-Intensity Profiles in Laser-Produced PlasmasPhysical Review Letters, 1988
- Point radiographic source characterizationJournal of Applied Physics, 1987
- Dynamics of laser-irradiated planar targets measured by x-ray spectroscopyPhysical Review A, 1987
- Density and temperature profiles within laser-produced plasmas in the classical-transport regimeThe Physics of Fluids, 1983