A study on image quality provided by a kilovoltage cone‐beam computed tomography
Authors:
Open Access
- 7 January 2013
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 14 (1) , 239-257
Abstract
The image‐guided radiotherapy technique (IGRT) makes use of imaging devices to verify the positions of the target volume and organs at risk during the treatment sessions. In this work we evaluate the image quality provided by an imaging system based on a kilovoltage cone‐beam CT, and explore its ability to perform IGRT and adaptive radiotherapy. We analyze the accuracy of the image slice width, the spatial resolution using the MTF function, the image uniformity, the signal‐to‐noise ratio, the contrast‐to‐noise ratio, the low‐contrast sensitivity, and the HU linearity with density. The studied parameters are evaluated in an objective and quantitative way, allowing for a direct comparison with other imaging devices. We conclude that the analyzed cone‐beam imaging system is adequate to accurately perform IGRT within its clinical use, despite the high level of noise present in a cone beam caused by scatter. We also point out the presence of a bowtie wobble artifact in the reconstructed images. Nevertheless, we conclude that these features do not limit the capability of the system to perform adaptive radiotherapy in most cases.PACS number: 87.57.‐sKeywords
This publication has 20 references indexed in Scilit:
- Does the planning dose–volume histogram represent treatment doses in image-guided prostate radiation therapy? Assessment with cone-beam computerised tomography scansRadiotherapy and Oncology, 2011
- Kilovoltage cone‐beam CT: Comparative dose and image quality evaluations in partial and full‐angle scan protocolsMedical Physics, 2010
- Quantitative evaluation of cone-beam computed tomography in target volume definition for offline image-guided radiation therapy of prostate cancerRadiotherapy and Oncology, 2009
- Dose calculation accuracy using cone-beam CT (CBCT) for pelvic adaptive radiotherapyPhysics in Medicine & Biology, 2009
- Cone beam computerized tomography: the effect of calibration of the Hounsfield unit number to electron density on dose calculation accuracy for adaptive radiation therapyPhysics in Medicine & Biology, 2009
- Adaptive Radiotherapy for Prostate Cancer Using Kilovoltage Cone-Beam Computed Tomography: First Clinical ResultsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- A study on adaptive IMRT treatment planning using kV cone-beam CTRadiotherapy and Oncology, 2007
- Online planning and delivery technique for radiotherapy of spinal metastases using cone-beam CT: Image quality and system performanceInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- A quality assurance program for the on‐boardMedical Physics, 2006
- Clinical use of electronic portal imaging: Report of AAPM Radiation Therapy Committee Task Group 58Medical Physics, 2001