Towards Effective and Efficient Patient-Specific Quality Assurance for Spot Scanning Proton Therapy
Authors:
- X. Ronald. ZhuXZX. Ronald. Zhu
- Yupeng LiYLYupeng Li
- Dennis MackinDMDennis Mackin
- Heng LiHLHeng Li
- Falk PoenischFPFalk Poenisch
- Andrew K. LeeALAndrew K. Lee
- Anita MahajanAMAnita Mahajan
- Steven J. FrankSFSteven J. Frank
- Michael T. GillinMGMichael T. Gillin
- Narayan SahooNSNarayan Sahoo
- Xiaodong ZhangXiaodong Zhang
Open Access
Abstract
An intensity-modulated proton therapy (IMPT) patient-specific quality assurance (PSQA) program based on measurement alone can be very time consuming due to the highly modulated dose distributions of IMPT fields. Incorporating independent dose calculation and treatment log file analysis could reduce the time required for measurements. In this article, we summarize our effort to develop an efficient and effective PSQA program that consists of three components: measurements, independent dose calculation, and analysis of patient-specific treatment delivery log files. Measurements included two-dimensional (2D) measurements using an ionization chamber array detector for each field delivered at the planned gantry angles with the electronic medical record (EMR) system in the QA mode and the accelerator control system (ACS) in the treatment mode, and additional measurements at depths for each field with the ACS in physics mode and without the EMR system. Dose distributions for each field in a water phantom were calculated independently using a recently developed in-house pencil beam algorithm and compared with those obtained using the treatment planning system (TPS). The treatment log file for each field was analyzed in terms of deviations in delivered spot positions from their planned positions using various statistical methods. Using this improved PSQA program, we were able to verify the integrity of the data transfer from the TPS to the EMR to the ACS, the dose calculation of the TPS, and the treatment delivery, including the dose delivered and spot positions. On the basis of this experience, we estimate that the in-room measurement time required for each complex IMPT case (e.g., a patient receiving bilateral IMPT for head and neck cancer) is less than 1 h using the improved PSQA program. Our experience demonstrates that it is possible to develop an efficient and effective PSQA program for IMPT with the equipment and resources available in the clinic.Keywords
This publication has 46 references indexed in Scilit:
- Efficient and reliable 3D dose quality assurance for IMRT by combining independent dose calculations with measurementsMedical Physics, 2013
- Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assuranceMedical Physics, 2013
- Water and tissue equivalence of a new PRESAGE® formulation for 3D proton beam dosimetry: A Monte Carlo studyMedical Physics, 2012
- Verification of proton range, position, and intensity in IMPT with a 3D liquid scintillator detector systemMedical Physics, 2012
- Parameterization of multiple Bragg curves for scanning proton beams using simultaneous fitting of multiple curvesPhysics in Medicine & Biology, 2011
- Dose reconstruction for volumetric modulated arc therapy (VMAT) using cone-beam CT and dynamic log filesPhysics in Medicine & Biology, 2010
- QA of intensity-modulated beams using dynamic MLC log filesJournal of Medical Physics, 2006
- Treatment planning and verification of proton therapy using spot scanning: Initial experiencesMedical Physics, 2004
- Verification of dynamic and segmental IMRT delivery by dynamic log file analysisJournal of Applied Clinical Medical Physics, 2002
- An analytical approximation of the Bragg curve for therapeutic proton beamsMedical Physics, 1997