Comparison of Dual Beam Dispersive and FTNIR Spectroscopy for Lactate Detection
Open Access
Abstract
Near Infrared (800–2500 nm) spectroscopy has been extensively used in biomedical applications, as it offers rapid, in vivo, bed-side monitoring of important haemodynamic parameters, which is especially important in critical care settings. However, the choice of NIR spectrometer needs to be investigated for biomedical applications, as both the dual beam dispersive spectrophotomer and the FTNIR spectrometer have their own advantages and disadvantages. In this study, predictive analysis of lactate concentrations in whole blood were undertaken using multivariate techniques on spectra obtained from the two spectrometer types simultaneously and results were compared. Results showed significant improvement in predicting analyte concentration when analysis was performed on full range spectral data. This is in comparison to analysis of limited spectral regions or lactate signature peaks, which yielded poorer prediction models. Furthermore, for the same region, FTNIR showed 10% better predictive capability than the dual beam dispersive NIR spectrometer.Keywords
Funding Information
- Engineering and Physical Sciences Research Council (EP/R003750/1)
This publication has 44 references indexed in Scilit:
- Determination of NIR informative wavebands for transmission non-invasive blood glucose measurement using a Fourier transform spectrometerAIP Advances, 2018
- A Brief Review of OPT101 Sensor Application in Near-Infrared Spectroscopy Instrumentation for Intensive Care Unit ClinicsSensors, 2017
- The Early Days of Commercial FT-IR Spectrometry: A Personal PerspectiveApplied Spectroscopy, 2017
- Identification of informative bands in the short-wavelength NIR region for non-invasive blood glucose measurementBiomedical Optics Express, 2016
- Near-Infrared SpectroscopySeminars in Cardiothoracic and Vascular Anesthesia, 2016
- Measurement of lactate in whole human blood with near-infrared transmission spectroscopyTalanta, 2003
- Designing the Ideal Process Analyser—Or at Least Making the Attempt (Continued)NIR News, 1994
- Nondestructive and Noninvasive Monitoring of Deoxyhemoglobin in the Vein by Use of a Near-Infrared Reflectance Spectrometer with a Fiber-Optic ProbeApplied Spectroscopy, 1992
- Noninvasive, Infrared Monitoring of Cerebral and Myocardial Oxygen Sufficiency and Circulatory ParametersScience, 1977
- THE INFRACORD® DOUBLE BEAM SPECTROPHOTOMETERChemical & Engineering News, 1957