Comparing Patellofemoral Kinematics Assessed With a Novel Muscle Actuator System and an Oxford Rig Using Noncadaveric Knees
- 17 January 2025
- journal article
- research article
- Published by Wiley in Journal of Biomechanical Engineering
- Vol. 147 (3) , 1-31
Abstract
Total knee replacement (TKR) failure, low patient satisfaction and high revision surgery rates may stem from insufficient preclinical testing. Conventional joint motion simulators for preclinical testing of TKR implants manipulate a knee joint in force, displacement, or simulated muscle control. However, a rig capable of using all three control modes has yet to be described in literature. This study aimed to validate a novel platform, the muscle actuator system (MAS), that can generate gravity-dependent, quadriceps-controlled squatting motions representative of an Oxford rig knee simulator and is mounted onto a force/displacement-control-capable joint motion simulator. Synthetic knee joint phantoms were created that comprised revision TKR implants and key extensor and flexor mechanism analogues, but no ligaments. The combined system implemented a constant force vector acting from simulated hip-to-ankle coordinates, effectively replicating gravity as observed in an Oxford rig. Quadriceps forces and patellofemoral joint kinematics were measured to assess the performance of the MAS and these tests showed high levels of repeatability and reproducibility. Forces and kinematics measured at a nominal patellar tendon length, and with patella alta and baja, were compared against those measured under the same conditions using a conventional Oxford rig, the Pennsylvania State Knee Simulator (PSKS). There was disagreement in absolute kinematics and muscle forces, but similar trends resulting from changing prosthesis design or patellar tendon length.Keywords
Funding Information
- Canada Foundation for Innovation (JELF 38141)
- Mitacs (IT17266)
- Natural Sciences and Engineering Research Council of Canada (Discovery Grant RGPIN-05693-2018)
- Ontario Ministry of Economic Development and Innovation (Early Research Award ER18-14-197)
This publication has 49 references indexed in Scilit:
- Patella height influences patellofemoral contact and kinematics following cruciate‐retaining total knee replacementJournal of Orthopaedic Research, 2022
- Condylar‐Stabilized TKR May Not Fully Compensate for PCL‐Deficiency: An In Vitro Cadaver StudyJournal of Orthopaedic Research, 2019
- Computational evaluation of TKR stability using feedback‐controlled compressive loadingJournal of Orthopaedic Research, 2018
- High Interspecimen Variability in Engagement of the Anterolateral Ligament: An In Vitro Cadaveric StudyClinical Orthopaedics and Related Research, 2017
- Patient Satisfaction after Total Knee Arthroplasty: Who is Satisfied and Who is Not?Clinical Orthopaedics and Related Research, 2010
- Factors Affecting the Impingement Angle of Fixed- and Mobile-Bearing Total Knee Replacements: A Laboratory StudyThe Journal of Arthroplasty, 2007
- Patellofemoral kinematics during knee flexion-extension: An in vitro studyJournal of Orthopaedic Research, 2006
- Standardisation of the description of patellofemoral motion and comparison between different techniquesKnee Surgery, Sports Traumatology, Arthroscopy, 2002
- Validation of a three-dimensional model of the kneeJournal of Biomechanics, 1996
- Articular contact in a three-dimensional model of the kneeJournal of Biomechanics, 1991