A novel high performance stopped-flow apparatus equipped with a special constructed mixing chamber containing a plunger under inert condition with a very short dead-time to investigate very rapid reactions
Open Access
- 1 November 2015
- journal article
- research article
- Published by Scientific Scholar in Arabian Journal of Chemistry
- Vol. 8 (6) , 873-884
Abstract
The present work set out to establish a novel stopped-flow instrument equipped with a special constructed mixing chamber containing a plunger to enable a kinetic study of the very rapid reactions under a dry inert atmosphere glove bag, in particular, for the reactions are sensitive to moisture or air. A stopped-flow spectrophotometer is essentially a conventional spectrophotometer with the addition of a system for rapid mixing of solutions. The purpose of this work is to describe the fabrication and evaluation of specially constructed and in-expensive stopped-flow system. The evaluation includes determination of the dead-time, relative mixing efficiency, and the measurement of known rate constants. Herein, a dead-time of about 3.4 ms was determined in the final modified construction of the stopped-flow apparatus in order to investigate the rapid initial during which some form of reaction intermediate is presented to be formed.Keywords
Funding Information
- Universities of Salford
- Sistan & Baluchestan
This publication has 20 references indexed in Scilit:
- Dielectric resonator-based stopped-flow electron paramagnetic resonanceReview of Scientific Instruments, 1994
- Simultaneous kinetic and spectral analysis with a vidicon rapid-scanning stopped-flow spectrometerAnalytical Chemistry, 1977
- Computer-interactive system for stopped-flour kinetics with rapid scanning molecular absorption spectrometryAnalytical Chemistry, 1975
- Variable temperature computerized dual-beam, rapid scanning stopped-flow apparatus for the study of air-sensitive systems and transients in enzyme reactionsAnalytical Chemistry, 1975
- Evaluation of a vidicon scanning spectrometer for ultraviolet molecular absorption spectrometryAnalytical Chemistry, 1975
- Computer controlled rapid scanning stopflow spectrometerAnalytical Chemistry, 1974
- A simple stopped-flow photometerAnalytical Biochemistry, 1969