Cardiac output measurements using the bioreactance technique in critically ill patients
Open Access
- 9 November 2012
- journal article
- letter
- Published by Springer Nature in Critical Care
- Vol. 16 (6) , 1-2
Abstract
Measurement of cardiac output (CO) using minimally invasive devices has gained popularity. In 11 patients we compared CO values obtained using the bioreactance technique - a new continuous, totally non-invasive CO monitor - with those obtained by semi-continuous thermodilution using a pulmonary artery catheter. We obtained CO measurements at study inclusion and after any relevant change in hemodynamic status (spontaneous or during fluid challenge, inotrope or vasopressor infusions). There was a poor correlation between the two techniques (r = 0.145). These data suggest that caution should be applied when using bioreactance devices in critically ill patients.Keywords
This publication has 5 references indexed in Scilit:
- Cardiac output measurements using the bioreactance technique in critically ill patientsCritical Care, 2012
- Fluid responsiveness predicted by noninvasive Bioreactance-based passive leg raise testIntensive Care Medicine, 2010
- Comparison between Flotrac-Vigileo and Bioreactance, a totally noninvasive method for cardiac output monitoringCritical Care, 2009
- Multicenter Evaluation of Noninvasive Cardiac Output Measurement by Bioreactance TechniqueJournal of Clinical Monitoring and Computing, 2008
- Noninvasive cardiac output monitoring (NICOM): a clinical validationIntensive Care Medicine, 2007