Towards MRI temperature mapping in real time—the proton resonance frequency method with undersampled radial MRI and nonlinear inverse reconstruction
Authors:
- 1 April 2017
- journal article
- research article
- Published by AME Publishing Company in Quantitative Imaging in Medicine and Surgery
- Vol. 7 (2) , 251-258
Abstract
Background: Optimal control of minimally invasive interventions by hyperthermia requires dynamic temperature mapping at high temporal resolution. Methods: Based on the temperature-dependent shift of the proton resonance frequency (PRF), this work developed a method for real-time MRI thermometry which relies on highly undersampled radial FLASH MRI sequences with iterative image reconstruction by regularized nonlinear inversion (NLINV). As a first step, the method was validated with use of a temperature phantom and ex vivo organs (swine kidney) subjected to heating by warm water or a pulsed laser source. Results: The temperature maps obtained by real-time PRF MRI demonstrate good accuracy as independently controlled by fiber-optic temperature sensors. Moreover, the dynamic results demonstrate both excellent sensitivity to single laser pulses (20 ms duration, 6 J energy output) and high temporal resolution, i.e., 200 ms acquisition times per temperature map corresponding to a rate of 5 frames per second. In addition, future extensions to in vivo applications were prepared by addressing the breathing-related motion problem by a pre-recorded library of reference images representative of all respiratory states. Conclusions: The proposed method for real-time MRI thermometry now warrants further developments towards in vivo MRI monitoring of thermal interventions in animals.Keywords
This publication has 28 references indexed in Scilit:
- Real‐time MRI of speaking at a resolution of 33 ms: Undersampled radial FLASH with nonlinear inverse reconstructionMagnetic Resonance in Medicine, 2012
- Combining two‐dimensional spatially selective RF excitation, parallel imaging, and UNFOLD for accelerated MR thermometry imagingMagnetic Resonance in Medicine, 2011
- MR thermometryJournal of Magnetic Resonance Imaging, 2008
- Technical aspects of lasers in urologyWorld Journal of Urology, 2007
- In vivo porcine liver radiofrequency ablation with simultaneous MR temperature imagingJournal of Magnetic Resonance Imaging, 2006
- Real time monitoring of laser‐induced thermal changes in cartilage in vitro by using snapshot FLASHMagnetic Resonance in Medicine, 1997
- A precise and fast temperature mapping using water proton chemical shiftMagnetic Resonance in Medicine, 1995
- MR temperature mapping of focused ultrasound surgeryMagnetic Resonance in Medicine, 1994
- Monitoring of laser and freezinginduced ablation in the liver with T1‐weighted MR imagingJournal of Magnetic Resonance Imaging, 1992
- Observation by MR Imaging of In Vivo Temperature Changes Induced by Radio Frequency HyperthermiaJournal of Computer Assisted Tomography, 1990