The use of exit detector sinograms to detect anatomical variations for patients extending beyond the TomoTherapy field of view: A feasibility study
Authors:
- H. Omar WootenHWH. Omar Wooten
- S. Murty GodduSGS. Murty Goddu
- Vivian RodriguezVRVivian Rodriguez
- Jeremy CatesJCJeremy Cates
- Perry GrigsbyPGPerry Grigsby
- Daniel A. LowDLDaniel A. Low
- 2 October 2012
- journal article
- Published by Wiley in Medical Physics
- Vol. 39 (10) , 6407-6419
Abstract
This work describes an independent method to use the TomoTherapy Hi-ART megavoltage CT imaging system for daily monitoring of anatomical changes of cancer patients whose anatomy extends beyond the imaging field of view. The imaging detector response to changes in attenuating media was measured using water-equivalent plastic. Weight loss was simulated using an anthropomorphic phantom and determining the system's ability to detect the weight loss. Layers of tissue-equivalent bolus were added to an anthropomorphic pelvis phantom and CT simulations of the phantom were conducted, one in which the phantom and bolus were both within the TomoTherapy imaging field of view, and another in which the couch was raised so that the bolus was outside the field of view. Gynecological treatment plans were developed using the TomoTherapy treatment planning system, and successive fractions of the plan were then delivered to the phantom. Weight loss was simulated by removing a 0.5 cm layer of bolus following each fraction. The exit detector sinograms were obtained from each fraction, and ratios of sinograms were calculated relative to a reference sinogram for which all bolus was in place. Histograms of ratio sinograms were determined and used to correlate with simulated weight loss. Exit detector sinograms and ratio histograms were also retrospectively analyzed for five patients all of whose anatomies extended beyond the imaging field of view and all of whom experienced weight variations exceeding 10% during treatment. Exit detector signal is well correlated to changes in attenuator thickness as demonstrated in both slab and anthropomorphic phantom geometries. Measured and expected signal increases agreed to within less than 2% for simulated weight loss on the anthropomorphic phantom. Exit detector signals for pelvic patients with significant weight loss variations were consistent with phantom measurements. The analysis of the ratio sinograms for the phantom measurements and real patients indicated that exit detector sinograms can be used to detect relative changes in patient anatomy for each fraction as a means of in vivo quality assurance.Keywords
This publication has 22 references indexed in Scilit:
- A diagnostic tool for basic daily quality assurance of a tomotherapy Hi•Art machineJournal of Applied Clinical Medical Physics, 2009
- Dosimetric effects of rotational output variation and x‐ray target degradation on helical tomotherapy plansMedical Physics, 2009
- A literature review of electronic portal imaging for radiotherapy dosimetryRadiotherapy and Oncology, 2008
- Dose calculation using megavoltage cone-beam CTInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2007
- Helical tomotherapy dynamic quality assuranceMedical Physics, 2006
- Deformable registration of the planning image (kVCT) and the daily images (MVCT) for adaptive radiation therapyPhysics in Medicine & Biology, 2006
- The use of megavoltage CT (MVCT) images for dose recomputationsPhysics in Medicine & Biology, 2005
- Accuracy of a wireless localization system for radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Monte Carlo study of a highly efficient gas ionization detector for megavoltage imaging and image-guided radiotherapyMedical Physics, 2002
- TomotherapySeminars in Radiation Oncology, 1999