Purpose: The objective was to prepare tantalum (Ta)-coated sandblasted and acid-etched (SLA) dental implants using vacuum plasma spraying (VPS) and to analyze their morphologies. Materials and Methods: Twelve SLA implants were coated with Ta using VPS. The topographies of the coatings and Ta/SLA surface interfaces were examined using scanning electron microscopy. The thickness at 4 locations for 6 Ta-coated and 6 uncoated SLA implants and pore sizes of the neck, central, and root areas of Ta-coated implants were measured. SPSS v20.0 was used for statistical analysis. Results: The Ta coatings were rough and consisted of pitted structures with various pore sizes; no cracks were observed. The Ta/SLA surface interface was tightly bonded. The 95% confidence interval of the Ta coating thickness was (114.0759, 129.3574). The maximal pore diameter was concentrated at 200 to 400 nm. Conclusion: SLA dental implants were successfully coated with Ta using VPS. The nanoporous structure of these implants may facilitate osseointegration compared with uncoated SLA implants.Keywords
This publication has 27 references indexed in Scilit:
- Porous Tantalum Coatings Prepared by Vacuum Plasma Spraying Enhance BMSCs Osteogenic Differentiation and Bone Regeneration In Vitro and In VivoPLOS ONE, 2013
- Histologic Retrieval Analysis of a Porous Tantalum Metal Implant in an Infected Primary Total Knee ArthroplastyThe Journal of Arthroplasty, 2012
- Clinical and microbiological characteristics of peri‐implantitis cases: a retrospective multicentre studyClinical Oral Implants Research, 2011
- Discrete calcium phosphate nanocrystalline deposition enhances osteoconduction on titanium‐based implant surfacesJournal of Biomedical Materials Research Part A, 2008
- Morphology, proliferation, and osteogenic differentiation of mesenchymal stem cells cultured on titanium, tantalum, and chromium surfacesJournal of Biomedical Materials Research Part A, 2007
- Experimental and clinical performance of porous tantalum in orthopedic surgeryBiomaterials, 2006
- Electrochemical corrosion and metal ion release from Co-Cr-Mo prosthesis with titanium plasma spray coatingBiomaterials, 2005
- Bone formation at titanium porous oxide (TiUnite™) oral implants in type IV boneClinical Oral Implants Research, 2004
- Influence of implant microstructure on the osseointegration of immediate implants placed in periodontally infected sitesClinical Oral Implants Research, 2004
- Biocompatibility and osteogenesis of refractory metal implants, titanium, hafnium, niobium, tantalum and rheniumBiomaterials, 2001