Measurement and effects of MOSKIN detectors on skin dose during high energy radiotherapy treatment
- 13 September 2012
- journal article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 35 (3) , 321-328
Abstract
During in vivo dosimetry for megavoltage X-ray beams, detectors such as diodes, Thermo luminescent dosimeters (TLD’s) and MOSFET devices are placed on the patient’s skin. This of course will affect the skin dose delivered during that fraction of the treatment. Whilst the overall impact on increasing skin dose would be minimal, little has been quantified concerning the level of increase in absorbed dose, in vivo dosimeters produce when placed in the beams path. To this extent, measurements have been made and analysis performed on dose changes caused by MOSKIN, MOSFET, skin dose detectors. Maximum increases in skin dose were measured as 15 % for 6 MV X-rays and 10 % for 10 MV X-rays at the active crystal of the MOSKIN device which is the thickest part of the detector. This is compared to 32 and 26 % for a standard 1 mm thick LiF TLD at 10 × 10 cm2 field size for 6 and 10 MV X-rays respectively. Radiochromic film, EBT2 has been shown to provide a high resolution 2 dimensional map of skin dose from these detectors and measures the effects of in vivo dosimeters used for radiotherapy dose assessment.Keywords
This publication has 30 references indexed in Scilit:
- Midline Dose Verification with Diode In Vivo Dosimetry for External Photon Therapy of Head and Neck and Pelvis Cancers During Initial Large-Field TreatmentsMedical Dosimetry, 2010
- Measuring solar UV radiation with EBT radiochromic filmPhysics in Medicine & Biology, 2010
- Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapyPhysics in Medicine & Biology, 2010
- A virtual source model of electron contamination of a therapeutic photon beamPhysics in Medicine & Biology, 2009
- Megavoltage x-ray skin dose variation with an angle using grid carbon fibre couch topsPhysics in Medicine & Biology, 2007
- Measurement of magnetic fields produced by a “Magnetic deflector” for the removal of electron contamination in radiotherapyAustralasian Physics & Engineering Sciences in Medicine, 2006
- Does mechanical pressure on radiochromic film affect optical absorption and dosimetry?Australasian Physics & Engineering Sciences in Medicine, 2006
- Post-irradiation colouration of Gafchromic EBT radiochromic filmPhysics in Medicine & Biology, 2005
- Radiochromic film for medical radiation dosimetryMaterials Science and Engineering: R: Reports, 2003
- TLD extrapolation for skin dose determination in vivoRadiotherapy and Oncology, 1996