A virtual source model of electron contamination of a therapeutic photon beam
Authors:
- M SikoraMSM Sikora
- M AlberMAM Alber
- 20 November 2009
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 54 (24) , 7329-7344
Abstract
The most efficient way of generating particles for Monte Carlo (MC) dose calculation is through a virtual source model (VSM) of the linear accelerator head. We have previously developed a VSM based on three sources: a primary photon source, a secondary photon source and an electron contamination source (Sikora et al 2007). In this work, we present an improvement of the electron contamination source. The VSM of contamination electrons (eVSM) is derived from a full MC simulation of the accelerator head with the BEAMnrc MC system. It comprises a Gaussian source located at the base of the flattening filter. The eVSM models two effects: an energy-dependent source diameter and an angular dependence of the particle fluence. The air scatter of the contamination electrons is approximated by energetic properties of the eVSM so that explicit in-air transport is not required during MC simulation of the dose distributions in the patient. The calculations of electron dose distributions were compared between the eVSM and the full MC simulation. Good agreement was achieved for various rectangular field sizes as well as for complex conformal segment shapes for the contamination electrons of 6 and 15 MV beams. The 3D dose evaluation of the surface dose in a CT-based patient geometry shows high accuracy (2%/2 mm) of the eVSM for both energies. The model has one tunable parameter, the mean energy of the spectrum at the patient surface. High accuracy and efficiency of particle generation make the eVSM a valuable virtual source of contamination electrons for MC treatment planning systems.Keywords
This publication has 24 references indexed in Scilit:
- Skin Dose Impact from Vacuum Immobilization Device and Carbon Fiber Couch in Intensity Modulated Radiation Therapy for Prostate CancerMedical Dosimetry, 2009
- Comparison between TG-51 and TRS-398: electron contamination effect on photon beam-quality specificationPhysics in Medicine & Biology, 2003
- A revision of the -evaluation concept for the comparison of dose distributionsPhysics in Medicine & Biology, 2003
- A virtual photon energy fluence model for Monte Carlo dose calculationMedical Physics, 2003
- Energy spectra, angular spread, fluence profiles and dose distributions of 6 and 18 MV photon beams: results of Monte Carlo simulations for a Varian 2100EX acceleratorPhysics in Medicine & Biology, 2002
- A multiple source model for 6 MV photon beam dose calculations using Monte CarloPhysics in Medicine & Biology, 2001
- Photon beam characterization and modelling for Monte Carlo treatment planningPhysics in Medicine & Biology, 2000
- Fast Monte Carlo dose calculation for photon beams based on the VMC electron algorithmMedical Physics, 1999
- An MCNP-based model of a linear accelerator x-ray beamPhysics in Medicine & Biology, 1999
- Magnetic analysis of the radiation components of a 10 MV photon beamMedical Physics, 1982