Evaluation of standardization capability of current cardiac troponin I assays by a correlation study: results of an IFCC pilot project
Authors:
- Jillian R. TateJTJillian R. Tate
- Mauro PanteghiniMPMauro Panteghini
- for the IFCC Working Group on StandFSfor the IFCC Working Group on Stand
- 1 January 2015
- journal article
- research article
- Published by Walter de Gruyter GmbH in Clinical Chemistry and Laboratory Medicine (CCLM)
- Vol. 53 (5) , 677-690
Abstract
Background: As a part of an International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) project to prepare a commutable reference material for cardiac troponin I (cTnI), a pilot study evaluated current cTnI assays for measurement equivalence and their standardization capability. Methods: cTnI-positive samples collected from 90 patients with suspected acute myocardial infarction were assessed for method comparison by 16 cTnI commercial assays according to predefined testing protocols. Seven serum pools prepared from these samples were also assessed. Results: Each assay was assessed against median cTnI concentrations measured by 16 cTnI assays using Passing-Bablok regression analysis of 79 patient samples with values above each assay’s declared detection limit. We observed a 10-fold difference in cTnI concentrations for lowest to highest measurement results. After mathematical recalibration of assays, the between-assay variation for patient samples reduced on average from 40% to 22% at low cTnI concentration, 37%–20% at medium concentration, and 29%–14% at high concentration. The average reduction for pools was larger at 16%, 13% and 7% for low, medium and high cTnI concentrations, respectively. Overall, assays demonstrated negligible bias after recalibration (y-intercept: –1.4 to 0.3 ng/L); however, a few samples showed substantial positive and/or negative differences for individual cTnI assays. Conclusions: All of the 16 commercial cTnI assays evaluated in the study demonstrated a significantly higher degree of measurement equivalence after mathematical recalibration, indicating that measurement harmonization or standardization would be effective at reducing inter-assay bias. Pooled sera behaved similarly to individual samples in most assays.Keywords
This publication has 10 references indexed in Scilit:
- Concordance, Variance, and Outliers in 4 Contemporary Cardiac Troponin Assays: Implications for HarmonizationClinical Chemistry, 2012
- Circulating Cardiac Troponin‐I Autoantibodies in Human Plasma and SerumAnnals of the New York Academy of Sciences, 2009
- Autoantibodies to Cardiac Troponin Associate with Higher Initial Concentrations and Longer Release of Troponin I in Acute Coronary Syndrome PatientsClinical Chemistry, 2009
- Toward Standardization of Cardiac Troponin I Measurements Part II: Assessing Commutability of Candidate Reference Materials and Harmonization of Cardiac Troponin I AssaysClinical Chemistry, 2006
- Standardization of Cardiac Troponin I Measurements: The Way Forward?Clinical Chemistry, 2005
- Selection of Antibodies and Epitopes for Cardiac Troponin Immunoassays: Should We Revise Our Evidence-Based Beliefs?Clinical Chemistry, 2005
- Comparison of Cardiac Troponin I Immunoassays Variably Affected by Circulating AutoantibodiesClinical Chemistry, 2005
- Negative Interference in Cardiac Troponin I Immunoassays by Circulating Troponin AutoantibodiesClinical Chemistry, 2005
- Current concepts in standardization of cardiac marker immunoassaysClinical Chemistry and Laboratory Medicine (CCLM), 2004
- Degradation of cardiac troponin I: implication for reliable immunodetectionClinical Chemistry, 1998