A low maintenance Sr optical lattice clock
Open Access
- 1 June 2016
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 723 (1) , 012019
Abstract
We describe the Sr optical lattice clock apparatus at NPL with particular emphasis on techniques used to increase reliability and minimise the human requirement in its operation. Central to this is a clock-referenced transfer cavity scheme for the stabilisation of cooling and trapping lasers. We highlight several measures to increase the reliability of the clock with a view towards the realisation of an optical timescale. The clock contributed 502 hours of data over a 25 day period (84 % uptime) in a recent measurement campaign with several uninterrupted periods of more than 48 hours. An instability of 2 × 10-17 was reached after 10 5 s of averaging in an interleaved self-comparison of the clock.Keywords
All Related Versions
This publication has 22 references indexed in Scilit:
- Systematic evaluation of an atomic clock at 2 × 10−18 total uncertaintyNature Communications, 2015
- The87Sr optical frequency standard at PTBMetrologia, 2011
- Precision Measurement of Gravity with Cold Atoms in an Optical Lattice and Comparison with a Classical GravimeterPhysical Review Letters, 2011
- Prospects for a Millihertz-Linewidth LaserPhysical Review Letters, 2009
- Measuring the frequency of a Sr optical lattice clock using a 120 km coherent optical transferOptics Letters, 2009
- Prospects for application of ultracoldmolecules in precision measurementsPhysical Review A, 2009
- Quantum Computing with Alkaline-Earth-Metal AtomsPhysical Review Letters, 2008
- The absolute frequency of the87Sr optical clock transitionMetrologia, 2008
- An optical lattice clock with spin-polarized 87Sr atomsThe European Physical Journal D, 2007
- SPECTROSCOPY OF STRONTIUM ATOMS IN THE LAMB-DICKE CONFINEMENT2002