Simulated shunts as dc low-resistance travelling standards
- 1 September 2015
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- p. 302-305
Abstract
We introduce a device that simulates a low-valued resistance standard (shunt) that can be employed as a metrology-grade travelling standard, for the traceability transfer of dc resistance in the low range, and in laboratory intercomparisons. The device is composed of a direct-current current comparator (DCCT) and a mid-range resistance standard. A prototype having a nominal resistance of 10m Ω has been constructed. Experiments show that it can be measured with high-accuracy commercial dc resistance bridges in their default operating configurations. The power coefficient is much lower than a corresponding shunt of the same nominal resistance. The expected long-term and transport stability is a few parts in 10 6 or better.Keywords
This publication has 10 references indexed in Scilit:
- Final report on bilateral comparison SIM.EM-S9.b: Comparison of resistance standards at 1 Ω and 10 kΩ between INIMET (Cuba) and INTI (Argentina), March to October 2012Metrologia, 2014
- Low-Ohmic Resistance Comparison: Measurement Capabilities and Resistor Traveling BehaviorIEEE Transactions on Instrumentation and Measurement, 2012
- Bilateral comparison of 1 Ω standards (ongoing BIPM key comparison BIPM.EM-K13.a) between the NSAI–NML (Ireland) and the BIPMMetrologia, 2012
- Electric current sensors: a reviewMeasurement Science and Technology, 2010
- Characteristics of precision 1 Ω standard resistors influencing transport behaviour and the uncertainty of key comparisonsMetrologia, 2009
- Bilateral comparison of 1 Ω standards (ongoing BIPM key comparison BIPM.EM-K13.a) between the NMIA (Australia) and the BIPMMetrologia, 2009
- Final report on the bilateral comparison of 1 Ω standards (ongoing BIPM key comparison BIPM.EM-K13.a) between the NIST (USA) and the BIPMMetrologia, 2008
- Behavior of resistors and shunts: with today's high-precision measurement capability and a century of materials experience, what can go wrong?IEEE Transactions on Instrumentation and Measurement, 1999
- Report on the 1990 International Comparison of 1 Ω and 10 kΩ Resistance Standards at the BIPMMetrologia, 1992
- A Direct-Current-Comparator Ratio Bridge for Four-Terminal Resistance MeasurementsIEEE Transactions on Instrumentation and Measurement, 1966