Hybrid pixel detectors are photon-counting devices for X-ray imaging and diffraction techniques at the fourth-generation Sirius synchrotron light source (LNLS/CNPEM). This work presents a comprehensive characterization regarding Modulation Transfer Function and pixel count rate linearity under high-flux conditions of the large-area PIMEGA detectors of 300 and 675 μm thick silicon sensors. They consist of photon-counting devices with a 55 microns pixel size based on Medipix3RX ASIC. The results validate the successful construction and performance of the Medipix3RX-based PIMEGA detectors. The silicon sensor platform offers a viable compromise between absorption efficiency and spatial resolution, making it a highly attractive technological solution for imaging and diffraction applications across the tender to initial hard X-ray energy ranges.