Tabulated square-shaped source model for linear accelerator electron beam simulation
- 1 January 2017
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cancer Research and Therapeutics
- Vol. 13 (1) , 69-79
Abstract
Context: Using this source model, the Monte Carlo (MC) computation becomes much faster for electron beams. Aims: The aim of this study was to present a source model that makes linear accelerator (LINAC) electron beam geometry simulation less complex. Settings and Design: In this study, a tabulated square-shaped source with transversal and axial distribution biasing and semi-Gaussian spectrum was investigated. Subjects and Methods: A low energy photon spectrum was added to the semi-Gaussian beam to correct the bremsstrahlung X-ray contamination. After running the MC code multiple times and optimizing all spectrums for four electron energies in three different medical LINACs (Elekta, Siemens, and Varian), the characteristics of a beam passing through a 10 cm × 10 cm applicator were obtained. The percentage depth dose and dose profiles at two different depths were measured and simulated. Results: The maximum difference between simulated and measured percentage of depth doses and dose profiles was 1.8% and 4%, respectively. The low energy electron and photon spectrum and the Gaussian spectrum peak energy and associated full width at half of maximum and transversal distribution weightings were obtained for each electron beam. The proposed method yielded a maximum computation time 702 times faster than a complete head simulation. Conclusions: Our study demonstrates that there was an excellent agreement between the results of our proposed model and measured data; furthermore, an optimum calculation speed was achieved because there was no need to define geometry and materials in the LINAC head.Keywords
This publication has 24 references indexed in Scilit:
- Simulation of dark current and dark current-induced background photons in the Thomson scattering X-ray sourceNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2015
- Simultaneous production of mixed electron–photon beam in a medical LINAC: A feasibility studyPhysica Medica, 2015
- Monte Carlo simulation and measurement of radiation leakage from applicators used in external electron radiotherapyPhysica Medica, 2013
- Monte Carlo investigation of the effect of small cutouts on beam profile parameters of 12 and 14 MeV electron beamsRadiation Measurements, 2013
- Determining optimization of the initial parameters in Monte Carlo simulation for linear accelerator radiotherapyRadiation Physics and Chemistry, 2013
- Simulations of a LINAC-based photoneutron sourceNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2012
- Evaluation of PENFAST – A fast Monte Carlo code for dose calculations in photon and electron radiotherapy treatment planningPhysica Medica, 2010
- Evaluation of a commercial VMC++ Monte Carlo based treatment planning system for electron beams using EGSnrc/BEAMnrc simulations and measurementsPhysica Medica, 2009
- Monte Carlo treatment planning for photon and electron beamsRadiation Physics and Chemistry, 2007
- Modeling and optimization of electron linac exit systems for nuclear technologiesNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006