Truncation Artifact on PET/CT: Impact on Measurements of Activity Concentration and Assessment of a Correction Algorithm

Authors:

Abstract
OBJECTIVE. Discrepancy between fields of view (FOVs) in a PET/CT scanner causes a truncation artifact when imaging extends beyond the CT FOV. The purposes of this study were to evaluate the impact of this artifact on measurements of 18F-FDG activity concentrations and to assess a truncation correction algorithm.MATERIALS AND METHODS. Two phantoms and five patients were used in this study. In the first phantom, three inserts (water, air, bone equivalent) were placed in a water-filled cylinder containing 18F-FDG. In the second phantom study, a chest phantom and a 2-L bottle fitted with a bone insert were used to simulate a patient's torso and arm. Both phantoms were imaged while positioned centrally (baseline) and at the edge of the CT FOV to induce truncation. PET images were reconstructed using attenuation maps from truncated and truncation-corrected CT images. Regions of interest (ROIs) drawn on the inserts, simulated arm, and background water of the baseline truncated and truncation-corrected PET images...