Abstract
A micropump‐actuated negative pressure pinched injection method is developed for parallel electrophoresis on a multi‐channel LIF detection system. The system has a home‐made device that could individually control 16‐port solenoid valves and a high‐voltage power supply. The laser beam is excitated and distributes to the array separation channels for detection. The hybrid Glass‐PDMS microfluidic chip comprises two common reservoirs, four separation channels coupled to their respective pneumatic micropumps and two reference channels. Due to use of pressure as a driving force, the proposed method has no sample bias effect for separation. There is only one high‐voltage supply needed for separation without relying on the number of channels, which is significant for high‐throughput analysis, and the time for sample loading is shortened to 1 s. In addition, the integrated micropumps can provide the versatile interface for coupling with other function units to satisfy the complicated demands. The performance is verified by separation of DNA marker and Hepatitis B virus DNA samples. And this method is also expected to show the potential throughput for the DNA analysis in the field of disease diagnosis.