Examination of blood haemoglobin concentration measured using the OSM2
- 1 January 2009
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 69 (1) , 92-101
Abstract
Blood haemoglobin concentration is regularly measured automatically by instruments reporting the value in around 1 min. The OSM2 from Radiometer is an example. Results from this instrument have been compared with those of a reference method using the hemiglobincyanide principle. Four healthy, moderately trained, young men (non‐smokers) cycled for 2 min to exhaustion. Blood samples were drawn from indwelling catheters in the femoral artery and vein before exercise, during exercise and in the 1 h recovery. Blood haemoglobin concentration was analysed using both methods. The results of the OSM2 were linearly related to those of the control method, with a random variation of 0.14 mmol L−1 (1.5 %). For arterial blood, the OSM2 showed a systematic bias of −0.36 mmol L−1 (−4 %). For femoral venous samples the bias varied depending on the haemoglobin concentration, being negative at low concentrations and positive at high values (−3 to +2 %). Consequently, the arteriovenous (a‐v) difference differed systematically between the two methods. The varying bias in the results of the OSM2 for femoral‐venous samples correlated with pH, pCO2, O2 saturation of haemoglobin (sO2) and with the haemoglobin concentration itself (cHb). Partial correlation analyses suggest that only the latter two correlations were independent, while correlations of the bias with pH and pCO2 were removed when correcting for the effect of sO2 and cHb. In conclusion, the OSM2 measures the blood haemoglobin concentration fairly precisely, but there is a variable bias of up to 4 % in absolute value. Finally, the instrument does not report a‐v differences reliably.Keywords
This publication has 33 references indexed in Scilit:
- Arterio‐venous differences of blood acid–base status and plasma sodium caused by intense bicyclingActa Physiologica Scandinavica, 2000
- Near-patient measurements of methemoglobin, oxygen saturation, and total hemoglobinCritical Care Medicine, 1995
- Evaluation of two oximeters for use in cardiac catheterization laboratoriesCatheterization and Cardiovascular Diagnosis, 1993
- The clinical value of the haemoglobin concentration analysis (g/L) and the calculation of oxyhaemoglobin saturation (%) in an automatic blood gas analyser (ABL2, Radiometer A/S, Denmark)Scandinavian Journal of Clinical and Laboratory Investigation, 1990
- Accurate measurements of hemoglobin oxygen saturation, and fractions of carboxyhemoglobin and methemoglobin in fetal blood using Radiometer OSM3: Corrections for fetal hemoglobin fraction and pHScandinavian Journal of Clinical and Laboratory Investigation, 1990
- Simultaneous measurements of blood pH,pCO2,pO2and concentrations of hemoglobin and its derivates-A multicenter studyScandinavian Journal of Clinical and Laboratory Investigation, 1990
- The evaluation of a photometer (the Radiometer OSM2) for the determination of haemoglobin concentration and per cent oxyhaemoglobin and carboxyhaemoglobin in bloodJournal of Medical Engineering & Technology, 1981
- Evaluation of the OSM2 HemoximeterCardiovascular Research, 1978
- Hemoglobin pigments. spectrophotometric determination of oxy-, carboxy-, met-, and sulfhemoglobin in capillary bloodClinica Chimica Acta, 1972
- Hemoglobin pigments. Photometer for oxygen saturation, carboxyhemoglobin, and methemoglobin in capillary bloodClinica Chimica Acta, 1972