Current distribution of distributed all-polar cochlear implant stimulation mode measured in-situ
Open Access
- 31 October 2022
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 17 (10) , e0275961
Abstract
Oticon Medical cochlear implants use a stimulation mode called Distributed All-Polar (DAP) that connects all non-stimulating available intracochlear electrodes and an extracochlear reference electrode. It results in a complex distribution of current that is yet undescribed. The present study aims at providing a first characterization of this current distribution. A Neuro Zti was modified to allow the measurement of current returning to each electrode during a DAP stimulation and was implanted in an ex-vivo human head. Maps of distributed current were then created for different stimulation conditions with different charge levels. Results show that, on average, about 20% of current returns to the extracochlear reference electrode, while the remaining 80% is distributed between intracochlear electrodes. The position of the stimulating electrode changed this ratio, and about 10% more current to the extracochlear return in case of the first 3 basal electrodes than for apical and mid position electrodes was observed. Increasing the charge level led to small but significant change in the ratio, and about 4% more current to the extracochlear return was measured when increasing the charge level from 11.7 to 70 nC. Further research is needed to show if DAP yields better speech understanding than other stimulation modes.Keywords
Funding Information
- Ministère de l'Enseignement supérieur, de la Recherche et de l'Innovation (CIFRE 2013/0855)
- Ministère de l'Enseignement supérieur, de la Recherche et de l'Innovation (CIFRE 2013/0855)
This publication has 41 references indexed in Scilit:
- Pre-, Per- and Postoperative Factors Affecting Performance of Postlinguistically Deaf Adults Using Cochlear Implants: A New Conceptual Model over TimePLOS ONE, 2012
- Reducing current spread using current focusing in cochlear implant usersHearing Research, 2012
- Cochlear-implant spatial selectivity with monopolar, bipolar and tripolar stimulationHearing Research, 2012
- Cochlear infrastructure for electrical hearingHearing Research, 2011
- Partial tripolar cochlear implant stimulation: Spread of excitation and forward masking in the inferior colliculusHearing Research, 2010
- Probing the Electrode-Neuron Interface With Focused Cochlear Implant StimulationTrends in Amplification, 2010
- Identifying Cochlear Implant Channels with Poor Electrode-Neuron Interface: Partial Tripolar, Single-Channel Thresholds and Psychophysical Tuning CurvesEar & Hearing, 2010
- Role of Electrode Placement as a Contributor to Variability in Cochlear Implant OutcomesOtology & Neurotology, 2008
- Cochlear implant electrode configuration effects on activation threshold and tonotopic selectivityHearing Research, 2008
- Noise Susceptibility of Cochlear Implant Users: The Role of Spectral Resolution and SmearingJournal of the Association for Research in Otolaryngology, 2005