55Mn-based fiducial markers for rapid and automated RF coil localization for hyperpolarized13C MRI
- 31 July 2020
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 85 (1) , 518-530
Abstract
Purpose To use fiducial markers containing manganese 55 to rapidly localize carbon 13 (C-13) RF coils for correcting images for B(1)variation. Methods Hollow high-density polyethylene spheres were filled with 3M sodium permanganate and affixed to a rectangular(13)C-tuned RF coil. The relative positions of the markers and coil conductors were mapped using CT. Marker positions were measured by MRI using a series of 1D projections and automated peak detection. Once the coil location was determined, coil sensitivity was estimated using a quasi-static calculation. Simulations were performed to determine the minimum number of projections required for robust localization. Phantom experiments were used to confirm the accuracy of marker localization as well as the calculated coil sensitivity. Finally, in vivo validation was performed using hyperpolarized(13)C pyruvate in a rat model. Results In simulations, our algorithm was accurate in determining marker positions when at least 6 projections were used (RMSE 1.4 +/- 0.9 mm). These estimates were verified in phantom experiments, where markers locations were determined with an RMS accuracy of 1.3 mm. A minimum SNR of 4 was required for automated detection to perform accurately. Computed coil sensitivity had a median error of 17% when taken over the entire measured area and 5.7% over a central region. In a rat, correction for nonuniform reception and flip angle was able to normalize the signals arising from asymmetrically positioned kidneys. Conclusion Manganese 55 fiducial markers are an inexpensive and reliable method for rapidly localizing(13)C RF coils and correcting(13)C images for B(1)variation without user intervention.Keywords
Funding Information
- National Institute of Diabetes and Digestive and Kidney Diseases (5R01DK115987)
- National Institute of Biomedical Imaging and Bioengineering (P41EB013598)
This publication has 17 references indexed in Scilit:
- Fast and accurate localization of multiple RF markers for tracking in MRI-guided interventionsMagnetic Resonance Materials in Physics, Biology and Medicine, 2014
- Propagation of calibration errors in prospective motion correction using external trackingMagnetic Resonance in Medicine, 2013
- Metabolic Imaging of Patients with Prostate Cancer Using Hyperpolarized [1- 13 C]PyruvateScience Translational Medicine, 2013
- Combined parallel and partial fourier MR reconstruction for accelerated 8‐channel hyperpolarized carbon‐13 in vivo magnetic resonance Spectroscopic imaging (MRSI)Journal of Magnetic Resonance Imaging, 2013
- Multi-channel metabolic imaging, with SENSE reconstruction, of hyperpolarized [1-13C] pyruvate in a live rat at 3.0tesla on a clinical MR scannerJournal of Magnetic Resonance, 2011
- Accelerated spectroscopic imaging of hyperpolarized C‐13 pyruvate using SENSE parallel imagingNMR in Biomedicine, 2009
- Increase in signal-to-noise ratio of > 10,000 times in liquid-state NMRProceedings of the National Academy of Sciences, 2003
- Generalized ReciprocityJournal of Magnetic Resonance, 1998
- Imaging of the active B1 field in vivoMagnetic Resonance in Medicine, 1996
- The NMR phased arrayMagnetic Resonance in Medicine, 1990