Proper Management of the Clinical Exposure Index Based on Body Thickness Using Dose Optimization Tools in Digital Chest Radiography: A Phantom Study
Open Access
- 13 May 2021
- journal article
- research article
- Published by MDPI AG in International Journal of Environmental Research and Public Health
- Vol. 18 (10) , 5203
Abstract
In radiography, the exposure index (EI), as per the International Electrotechnical Commission standard, depends on the incident beam quality and exposure dose to the digital radiography system. Today automatic exposure control (AEC) systems are commonly employed to obtain the optimal image quality. An AEC system can maintain a constant incident exposure dose on the image receptor regardless of the patient thickness. In this study, we investigated the relationship between body thickness, entrance surface dose (ESD), EI, and the exposure indicator (S value) with the aim of using EI as the dose optimization tool in digital chest radiography (posterior–anterior and lateral projection). The exposure condition from the Korean national survey for determining diagnostic reference levels and two digital radiography systems (photostimulable phosphor plate and indirect flat panel detector) were used. As a result, ESD increased as the phantom became thicker with constant exposure indicator, which indicates similar settings to an AEC system, but the EI indicated comparatively constant values without following the tendency of ESD. Therefore, body thickness should be considered under the AEC system for introducing EI as the dose optimization tool in digital chest radiography.Keywords
Funding Information
- Korea Environment Institute (2020003070001)
This publication has 21 references indexed in Scilit:
- Effect of X-ray beam quality on determination of exposure indexRadiological Physics and Technology, 2015
- Evaluation of the effects of subject thickness on the exposure index in digital radiographyRadiological Physics and Technology, 2015
- Monte Carlo simulation-based feasibility study of a dose-area product meter built into a collimator for diagnostic X-rayRadiation Protection Dosimetry, 2013
- A comparative assessment of entrance surface doses in analogue and digital radiography during common radiographic examinationsRadiation Protection Dosimetry, 2013
- The standardized exposure index for digital radiography: an opportunity for optimization of radiation dose to the pediatric populationPediatric Radiology, 2011
- Quality assurance: using the exposure index and the deviation index to monitor radiation exposure for portable chest radiographs in neonatesPediatric Radiology, 2010
- Advances in Digital Radiography: Physical Principles and System OverviewRadioGraphics, 2007
- Optimisation in general radiographyBiomedical Imaging and Intervention Journal, 2007
- Diagnostic imaging over the last 50 years: research and development in medical imaging science and technologyPhysics in Medicine & Biology, 2006
- ICRP recommendations on ‘Managing patient dose in digital radiology’Radiation Protection Dosimetry, 2005