Implementation of synthetic aperture imaging on a hand-held device
- 1 September 2014
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 10510117,p. 2177-2180
Abstract
This paper presents several implementations of Synthetic Aperture Sequential Beamforming (SASB) on commercially available hand-held devices. The implementations include real-time wireless reception of ultrasound radio frequency signals and GPU processing for B-mode imaging. The proposed implementation demonstrates that SASB can be executed in-time for real-time ultrasound imaging. The wireless communication between probe and processing device satisfies the required bandwidth for real-time data transfer with current 802.11ac technology. The implementation is evaluated using four different hand-held devices all with different chipsets and a BK Medical UltraView 800 ultrasound scanner emulating a wireless probe. The wireless transmission is benchmarked using an imaging setup consisting of 269 scan lines × 1472 complex samples (1.58 MB pr. frame, 16 frames per second). The measured data throughput reached an average of 28.8 MB/s using a LG G2 mobile device, which is more than the required data throughput of 25.3 MB/s. Benchmarking the processing performance for B-mode imaging showed a total processing time of 18.9 ms (53 frames/s), which is less than the acquisition time (62.5 ms).Keywords
This publication has 7 references indexed in Scilit:
- Implementation of a versatile research data acquisition system using a commercially available medical ultrasound scannerIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 2012
- In Vivo Evaluation of Synthetic Aperture Sequential BeamformingUltrasound in Medicine & Biology, 2012
- Review article: Use of ultrasound in the developing worldInternational Journal of Emergency Medicine, 2011
- Medical ultrasound systemsInterface Focus, 2011
- Portable Ultrasound for Remote Environments, Part I: Feasibility of Field DeploymentThe Journal of Emergency Medicine, 2011
- Use of ultrasound by emergency medical services: a reviewInternational Journal of Emergency Medicine, 2008
- Image quality assessment: from error visibility to structural similarityIEEE Transactions on Image Processing, 2004