Tracking of rigid head motion duringMRIusing an EEGsystem
Open Access
- 25 April 2022
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 88 (2) , 986-1001
Abstract
Purpose: To demonstrate a novel method for tracking of head movements during MRI using electroencephalography (EEG) hardware for recording signals induced by native imaging gradients.Theory and Methods: Gradient switching during simultaneous EEG–fMRI induces distortions in EEG signals, which depend on subject head position and orientation. When EEG electrodes are interconnected with high‐impedance carbon wire loops, the induced voltages are linear combinations of the temporal gradient waveform derivatives. We introduce head tracking based on these signals (CapTrack) involving 3 steps: (1) phantom scanning is used to characterize the target sequence and a fast calibration sequence; (2) a linear relation between changes of induced signals and head pose is established using the calibration sequence; and (3) induced signals recorded during target sequence scanning are used for tracking and retrospective correction of head movement without prolonging the scan time of the target sequence. Performance of CapTrack is compared directly to interleaved navigators.Results: Head‐pose tracking at 27.5 Hz during echo planar imaging (EPI) was demonstrated with close resemblance to rigid body alignment (mean absolute difference: [0.14 0.38 0.15]‐mm translation, [0.30 0.27 0.22]‐degree rotation). Retrospective correction of 3D gradient‐echo imaging shows an increase of average edge strength of 12%/−0.39% for instructed/uninstructed motion with CapTrack pose estimates, with a tracking interval of 1561 ms and high similarity to interleaved navigator estimates (mean absolute difference: [0.13 0.33 0.12] mm, [0.28 0.15 0.22] degrees).Conclusion: Motion can be estimated from recordings of gradient switching with little or no sequence modification, optionally in real time at low computational burden and synchronized to image acquisition, using EEG equipment already found at many research institutions.Keywords
This publication has 47 references indexed in Scilit:
- Measurement and Correction of Microscopic Head Motion during Magnetic Resonance Imaging of the BrainPLOS ONE, 2012
- An embedded optical tracking system for motion-corrected magnetic resonance imaging at 7TMagnetic Resonance Materials in Physics, Biology and Medicine, 2012
- Prospective motion correction in brain imaging: A reviewMagnetic Resonance in Medicine, 2012
- PROMO: Real‐time prospective motion correction in MRI using image‐based trackingMagnetic Resonance in Medicine, 2009
- Prospective real‐time correction for arbitrary head motion using active markersMagnetic Resonance in Medicine, 2009
- Measurement and reduction of motion and ballistocardiogram artefacts from simultaneous EEG and fMRI recordingsNeuroImage, 2007
- Prospective Real-Time Slice-by-Slice Motion Correction for fMRI in Freely Moving SubjectsMagnetic Resonance Materials in Physics, Biology and Medicine, 2006
- Movement‐Related effects in fMRI time‐seriesMagnetic Resonance in Medicine, 1996
- Orbital navigator echoes for motion measurements in magnetic resonance imagingMagnetic Resonance in Medicine, 1995
- The Scree Test For The Number Of FactorsMultivariate Behavioral Research, 1966