Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation
Authors:
Open Access
- 18 April 2020
- journal article
- research article
- Published by Press of International Journal of Ophthalmology (IJO Press) in International Journal of Ophthalmology
- Vol. 13 (4) , 625-631
Abstract
To evaluate the intra-operator repeatability of time domain and swept-source Fourier domain anterior segment optical coherence tomography (AS-OCT), namely, Visante AS-OCT and Casia SS-1000 OCT, in measuring the preoperative parameters of implantable collamer lens (ICL) in myopic eyes, as well as the agreement between the two devices. A total of 97 eyes from 49 myopes were investigated in this prospective case series study. The anterior chamber depth (ACD), angle-to-angle distance (ATA), pupil diameter (PD) and crystalline lens rise (CLR) in all subjects were measured for three times during one session by the same operator. The repeatability was evaluated using the within-subject standard deviation (Sw), repeatability limits and intraclass correlation coefficients (ICC). The agreement between the two systems was evaluated using the Bland-Altman plots and 95% limits of agreement (LoA). The repeatability limits of Visante AS-OCT in measuring ACD, ATA, PD and CLR were 0.099, 0.141, 0.304, and 0.079 mm, respectively. The repeatability limits of Casia SS-1000 OCT in measuring ACD, ATA, PD, and CLR were 0.105, 0.127, 0.357, and 0.082 mm, respectively. Excellent repeatability could be attained in both devices, with the ICC>0.8 for all the measured variables. The interdevice agreement was excellent (P>0.05) for ACD and ATA, but poor (P<0.05) for PD and CLR. Good repeatability can be attained by time domain and swept-source Fourier-domain OCT for all the measured variables. Moreover, interdevice agreement analysis suggests that interchangeable measurements between two devices can be achieved for ACD and ATA, but not for PD and CLR; but the differences in measurements were not clinically significant.Keywords
This publication has 26 references indexed in Scilit:
- Implantable Collamer Lens Sizing Method Based on Swept-Source Anterior Segment Optical Coherence TomographyAmerican Journal of Ophthalmology, 2018
- Reproducibility and Agreement of Anterior Segment Parameter Measurements Obtained Using the CASIA2 and Spectralis OCT2 Optical Coherence Tomography DevicesJournal of Glaucoma, 2017
- Intrasession repeatability of ocular anatomical measurements obtained with a multidiagnostic device in healthy eyesBMC Ophthalmology, 2017
- Comparison of time-domain, spectral-domain and swept-source OCT in evaluating aqueous cells in vitroScience China Life Sciences, 2016
- Correcting Interdevice Bias of Horizontal White-to-White and Sulcus-to-Sulcus Measures Used for Implantable Collamer Lens SizingAmerican Journal of Ophthalmology, 2016
- Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmologyOphthalmic and Physiological Optics, 2011
- Anterior chamber angle imaging with optical coherence tomographyEye, 2011
- Repeatability, reproducibility, and agreement of central anterior chamber depth measurements in pseudophakic and phakic eyes: Optical coherence tomography versus ultrasound biomicroscopyJournal of Cataract & Refractive Surgery, 2010
- Direct Measurement of the Ciliary Sulcus Diameter by 35-Megahertz Ultrasound BiomicroscopyOphthalmology, 2007
- Assessment and reproducibility of anterior chamber depth measurement with anterior segment optical coherence tomography compared with immersion ultrasonographyJournal of Cataract & Refractive Surgery, 2007